Automated, continuous and non-invasive assessment of pulse pressure variations using CNAP® system

Archive ouverte

Biais, Matthieu | Stecken, Laurent | Martin, Aurélie | Roullet, Stéphanie | Quinart, Alice | Sztark, François

Edité par CCSD ; Springer Verlag -

International audience. Non-invasive respiratory variations in arterial pulse pressure using infrared-plethysmography (PPVCNAP) are able to predict fluid responsiveness in mechanically ventilated patients. However, they cannot be continuously monitored. The present study evaluated a new algorithm allowing continuous measurements of PPVCNAP (PPVCNAPauto) (CNSystem, Graz, Austria). Thirty-five patients undergoing vascular surgery were studied after induction of general anaesthesia. Stroke volume was measured using the VigileoTM/FloTracTM. Invasive pulse pressure variations were manually calculated using an arterial line (PPVART) and PPVCNAPauto was continuously displayed. PPVART and PPVCNAPauto were simultaneously recorded before and after volume expansion (500 ml hydroxyethylstarch). Subjects were defined as responders if stroke volume increased by ≥15 %. Twenty-one patients were responders. Before volume expansion, PPVART and PPVCNAPauto exhibited a bias of 0.1 % and limits of agreement from -7.9 % to 7.9 %. After volume expansion, PPVART and PPVCNAPauto exhibited a bias of -0.4 % and limits of agreement from -5.3 % to 4.5 %. A 14 % baseline PPVART threshold discriminated responders with a sensitivity of 86 % (95 % CI 64-97 %) and a specificity of 100 % (95 % CI 77-100 %). Area under the receiver operating characteristic (ROC) curve for PPVART was 0.93 (95 % CI 0.79-0.99). A 15 % baseline PPVCNAPauto threshold discriminated responders with a sensitivity of 76% (95 % CI 53-92 %) and a specificity of 93 % (95 % CI 66-99 %). Area under the ROC curves for PPVCNAPauto was 0.91 (95 % CI 0.76-0.98), which was not different from that for PPVART. When compared with PPVART, PPVCNAPauto performs satisfactorily in assessing fluid responsiveness in hemodynamically stable surgical patients.

Consulter en ligne

Suggestions

Du même auteur

Impact of Systemic Vascular Resistance on the Accuracy of the Pulsioflex Device

Archive ouverte | Biais, Matthieu | CCSD

Comment in Questionable Design to Validate the ProAQT/Pulsioflex Device. [Anesth Analg. 2017]. International audience. BACKGROUND:The accuracy of currently available devices using pulse contour analysis without exte...

Peri-operative management of pheochromocytoma with intravenous urapidil to prevent hemodynamic instability: A 17-year experience

Archive ouverte | Tauzin-Fin, Patrick | CCSD

International audience. Background and Aims: Surgery for pheochromocytoma (PCC) can cause excessive catecholamine release with severe hypertension. Alpha blockade is the mainstay of preoperative management. The aim ...

Dynamic arterial elastance obtained using arterial signal does not predict an increase in arterial pressure after a volume expansion in the operating room

Archive ouverte | Lanchon, Romain | CCSD

International audience. INTRODUCTION: Dynamic arterial elastance (Eadyn) is defined as the ratio between pulse pressure variations (PPV) and stroke volume variations (SVV). Eadyn has been proposed to predict an incr...