Impact of Systemic Vascular Resistance on the Accuracy of the Pulsioflex Device

Archive ouverte

Biais, Matthieu | Mazocky, Élodie | Stecken, Laurent | Pereira, Bruno | Sesay, Musa | Roullet, Stéphanie | Quinart, Alice | Sztark, François

Edité par CCSD ; Lippincott, Williams & Wilkins -

Comment in Questionable Design to Validate the ProAQT/Pulsioflex Device. [Anesth Analg. 2017]. International audience. BACKGROUND:The accuracy of currently available devices using pulse contour analysis without external calibration for cardiac index (CI) estimation is negatively impacted by hyperdynamic states, low systemic vascular resistance (SVR), and abrupt changes in SVR. The aim of this study was to evaluate the accuracy of a new device, the Pulsioflex (Pulsion Medical System), in patients undergoing liver transplantation.METHODS:Thirty consecutive patients scheduled for liver transplantation were included. CI was monitored using pulmonary arterial catheter (CI-PAC) and Pulsioflex (CI-Pulsio). Simultaneous CI measurements were made intraoperatively at 9 different stages of the procedure.RESULTS:Two hundred seventy pairs of measurements were analyzed. The median CI-Pulsio values (3.3; interquartile range, 2.8-3.8 L·min·m) were significantly different from the median CI-PAC (4.1; interquartile range, 3.1-5.0 L·min·m; P < .0001). Bland and Altman analysis showed a mean bias of 0.8 L·min·m and 95% limit of agreement from -2.5 to 4.1 L·min·m. Percentage error was 65% (95% confidence interval, 60%-71%). Considering the variations in CI between 2 stages, the comparison between changes in CI-PAC and changes in CI-Pulsio showed a mean bias of 0.1 L·min·m and 95% limit of agreement of -2.1 to 2.2 L·min·m. When excluding changes in CI <0.5 L·min·m (154 paired analyzed), the concordance rate was 62% (95% confidence interval, 54%-70%). The bias between CI-PAC and CI-Pulsio was negatively correlated with SVR (r = -0.67, P < .0001). The bias between changes in CI-PAC and changes in CI-Pulsio was also negatively correlated with changes in SVR (r = -0.52, P < .0001).CONCLUSIONS:In patients undergoing liver transplantation, Pulsioflex does not accurately estimate CI. Its accuracy is highly impacted by SVR, and it is not able to track changes in CI when large variations in SVR occur.

Consulter en ligne

Suggestions

Du même auteur

Automated, continuous and non-invasive assessment of pulse pressure variations using CNAP® system

Archive ouverte | Biais, Matthieu | CCSD

International audience. Non-invasive respiratory variations in arterial pulse pressure using infrared-plethysmography (PPVCNAP) are able to predict fluid responsiveness in mechanically ventilated patients. However, ...

Peri-operative management of pheochromocytoma with intravenous urapidil to prevent hemodynamic instability: A 17-year experience

Archive ouverte | Tauzin-Fin, Patrick | CCSD

International audience. Background and Aims: Surgery for pheochromocytoma (PCC) can cause excessive catecholamine release with severe hypertension. Alpha blockade is the mainstay of preoperative management. The aim ...

Dynamic arterial elastance obtained using arterial signal does not predict an increase in arterial pressure after a volume expansion in the operating room

Archive ouverte | Lanchon, Romain | CCSD

International audience. INTRODUCTION: Dynamic arterial elastance (Eadyn) is defined as the ratio between pulse pressure variations (PPV) and stroke volume variations (SVV). Eadyn has been proposed to predict an incr...

Chargement des enrichissements...