Characterization of MOSFET detectors for in vivo dosimetry in interventional radiology and for dose reconstruction in case of overexposure

Archive ouverte

Bassinet, C. | Huet, C. | Baumann, M. | Etard, C. | Réhel, J.-L. | Boisserie, G. | Debroas, J. | Aubert, B. | Clairand, I.

Edité par CCSD ; Lippincott, Williams & Wilkins -

International audience. As MOSFET (Metal Oxide Semiconductor Field Effect Transistor) detectors allow dose measurements in real time, the interest in these dosimeters is growing. The aim of this study was to investigate the dosimetric properties of commercially available TN-502RD-H MOSFET silicon detectors (Best Medical Canada, Ottawa, Canada) in order to use them for in vivo dosimetry in interventional radiology and for dose reconstruction in case of overexposure. Reproducibility of the measurements, dose rate dependence, and dose response of the MOSFET detectors have been studied with a Co source. Influence of the dose rate, frequency, and pulse duration on MOSFET responses has also been studied in pulsed x-ray fields. Finally, in order to validate the integrated dose given by MOSFET detectors, MOSFETs and TLDs (LiFMg,Cu,P) were fixed on an Alderson-Rando phantom in the conditions of an interventional neuroradiology procedure, and their responses have been compared. The results of this study show the suitability of MOSFET detectors for in vivo dosimetry in interventional radiology and for dose reconstruction in case of accident, provided a well-corrected energy dependence, a pulse duration equal to or higher than 10 ms, and an optimized contact between the detector and the skin of the patient are achieved. Copyright © 2013 Health Physics Society.

Consulter en ligne

Suggestions

Du même auteur

Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones

Archive ouverte | Bassinet, C. | CCSD

Erratum: Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones (Medical Physics (2013) 40 (071725)) ...

SESAME A TOOL FOR NUMERICAL DOSIMETRIC RECONSTRUCTION OF PATIENTS OVEREXPOSURES IN INTERVENTIONAL RADIOLOGY

Archive ouverte | Huet, C. | CCSD

International audience. Radiation overexposure accidents are rare but can have severe health consequences. Evaluating the dose received by the patient is a crucial step in the medical management. For that purpose, f...

A new single crystal diamond dosimeter for small beam: comparison with different commercial active detectors

Archive ouverte | Marsolat, Fanny | CCSD

International audience. Recent developments of new therapy techniques using small photon beams, such as stereotactic radiotherapy, require suitable detectors to determine the delivered dose with a high accuracy. The...

Chargement des enrichissements...