Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones

Archive ouverte

Bassinet, C. | Huet, C. | Derreumaux, S. | Brunet, G. | Chéa, M. | Baumann, M. | Lacornerie, T. | Gaudaire-Josset, S. | Trompier, F. | Roch, P. | Boisserie, G. | Clairand, I.

Edité par CCSD ; American Association of Physicists in Medicine -

Erratum: Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones (Medical Physics (2013) 40 (071725)) : 10.1118/1.4823794. International audience. Purpose The use of small photon fields is now an established practice in stereotactic radiosurgery and radiotherapy. However, due to a lack of lateral electron equilibrium and high dose gradients, it is difficult to accurately measure the dosimetric quantities required for the commissioning of such systems. Moreover, there is still no metrological dosimetric reference for this kind of beam today. In this context, the first objective of this work was to determine and to compare small fields output factors (OF) measured with different types of active detectors and passive dosimeters for three types of facilities a CyberKnife® system, a dedicated medical linear accelerator (Novalis) equipped with m3 microMLC and circular cones, and an adaptive medical linear accelerator (Clinac 2100) equipped with an additional m3 microMLC. The second one was to determine the kQclin,Q msrfclin,fmsr correction factors introduced in a recently proposed small field dosimetry formalism for different active detectors. Methods Small field sizes were defined either by microMLC down to 6 × 6 mm2 or by circular cones down to 4 mm in diameter. OF measurements were performed with several commercially available active detectors dedicated to measurements in small fields (high resolution diodes IBA SFD, Sun Nuclear EDGE, PTW 60016, PTW 60017; ionizing chambers PTW 31014 PinPoint chamber, PTW 31018 microLion liquid chamber, and PTW 60003 natural diamond). Two types of passive dosimeters were used LiF microcubes and EBT2 radiochromic films. Results Significant differences between the results obtained by several dosimetric systems were observed, particularly for the smallest field size for which the difference in the measured OF reaches more than 20%. For passive dosimeters, an excellent agreement was observed (better than 2%) between EBT2 and LiF microcubes for all OF measurements. Moreover, it has been shown that these passive dosimeters do not require correction factors and can then be used as reference dosimeters. Correction factors for the active detectors have then been determined from the mean experimental OF measured by the passive dosimeters. Conclusions Four sets of correction factors needed to apply the new small field dosimetry formalism are provided for several active detectors. A protocol for small photon beams OF determination based on passive dosimeters measurements has been recently proposed to French radiotherapy treatment centers. © 2013 American Association of Physicists in Medicine.

Consulter en ligne

Suggestions

Du même auteur

Characterization of MOSFET detectors for in vivo dosimetry in interventional radiology and for dose reconstruction in case of overexposure

Archive ouverte | Bassinet, C. | CCSD

International audience. As MOSFET (Metal Oxide Semiconductor Field Effect Transistor) detectors allow dose measurements in real time, the interest in these dosimeters is growing. The aim of this study was to investi...

A new single crystal diamond dosimeter for small beam: comparison with different commercial active detectors

Archive ouverte | Marsolat, Fanny | CCSD

International audience. Recent developments of new therapy techniques using small photon beams, such as stereotactic radiotherapy, require suitable detectors to determine the delivered dose with a high accuracy. The...

Characterization and optimization of EBT2 radiochromic films dosimetry system for precise measurements of output factors in small fields used in radiotherapy

Archive ouverte | Huet, C. | CCSD

International audience. The accurate determination of absorbed dose in small photon beams, especially for stereotactic radiation therapy, is a difficult task with commercially available detectors. As these small fie...

Chargement des enrichissements...