Patient-Specific Finite-Element Simulation of the Insertion of Guidewire During an EVAR Procedure: Guidewire Position Prediction Validation on 28 Cases

Archive ouverte

Gindre, Juliette | Bel-Brunon, Aline | Lucas, Antoine | Rochette, Michel | Kaladji, Adrien | Haigron, Pascal | Combescure, Alain

Edité par CCSD ; Institute of Electrical and Electronics Engineers -

International audience. Objective: Validation of a numerical method to compute arterial deformations under the insertion of an “extra-siff” guidewire during Endovascular Repair of Abdominal Aortic Aneurysm. Methods: We propose the validation of a previously developed simulation method. The model is calibrated using anatomical hypothesis and intraoperative observations. Simulation results are blindly evaluated against 3-D imaging data acquired during the surgical procedure on 28 patients, based on the predicted position of the intraoperative guidewire. Results: Simulation was successfully conducted on the 28 patients. The mean position error given by the Modified Hausdorff Distance for the 28 cases was 3.8 ± 1.9 mm, which demonstrates very good results for most of the cases. Conclusion: The work reported here shows that numerical simulation can predict some rather large variations in the vascular geometry due to tools insertion, for a wide variety of aorto-iliac morphologies. This is a new step toward clinically applicable mechanical simulation. Significance: Validation on 3-D intraoperative data on a large number of cases - robustness on adverse anatomies.

Consulter en ligne

Suggestions

Du même auteur

Finite element simulation of the insertion of guidewires during an EVAR procedure: example of a complex patient case, a first step toward patient-specific parameterized models

Archive ouverte | Gindre, Juliette | CCSD

International audience. Deformations of the vascular structure due to the insertion of tools during endovascular treatment of aneurysms of the abdominal aorta, unless properly anticipated during the preoperative pla...

Influence of balloon design, plaque material composition, and balloon sizing on acute post angioplasty outcomes: An implicit finite element analysis

Archive ouverte | Helou, Bernard | CCSD

International audience. In this work we propose a generic modeling approach for simulating percutaneous transluminal angioplasty (PTA) endovascular treatment, and evaluating the influence of balloon design, plaque c...

Patient-specific finite element simulation of peripheral artery percutaneous transluminal angioplasty to evaluate the procedure outcome without stent implantation

Archive ouverte | Helou, Bernard | CCSD

International audience. The purpose of this work is to present a patient-specific (PS) modeling approach for simulating percutaneous transluminal angioplasty (PTA) endovascular treatment and assessing the balloon si...

Chargement des enrichissements...