Patient-specific finite element simulation of peripheral artery percutaneous transluminal angioplasty to evaluate the procedure outcome without stent implantation

Archive ouverte

Helou, Bernard | Bel-Brunon, Aline | Dupont, Claire | Ye, Wenfeng | Silvestro, Claudio | Rochette, Michel | Lucas, Antoine | Kaladji, Adrien | Haigron, Pascal

Edité par CCSD ; John Wiley and Sons -

International audience. The purpose of this work is to present a patient-specific (PS) modeling approach for simulating percutaneous transluminal angioplasty (PTA) endovascular treatment and assessing the balloon sizing influence on short-term outcomes in peripheral arteries, i.e. without stent implantation. Two 3D PS stenosed femoral artery models, one with a dominant calcified atherosclerosis while the other with a lipidic plaque, were generated from pre-operative computed tomography angiography images. Elastoplastic constitutive laws were implemented within the plaque and artery models. Implicit finite element method (FEM) was used to simulate the balloon inflation and deflation for different sizings. Besides vessel strains, results were mainly evaluated in terms of the elastic recoil ratio (ERR) and lumen gain ratio (LGR) attained immediately after PTA. Higher LGR values were shown within the stenosed region of the lipidic patient. Simulated results also showed a direct and quantified correlation between balloon sizing and LGR and ERR for both patients after PTA, with a more significant influence on the lumen gain. The max principal strain values in the outer arterial wall increased at higher balloon sizes during inflation as well, with higher rates of increase when the plaque was calcified. Results show that our model could serve in finding a compromise for each stenosis type: maximizing the achieved lumen gain after PTA, but at the same time without damaging the arterial tissue. The proposed methodology can serve as a step toward a clinical decision support system to improve angioplasty balloon sizing selection prior to the surgery.

Suggestions

Du même auteur

Influence of balloon design, plaque material composition, and balloon sizing on acute post angioplasty outcomes: An implicit finite element analysis

Archive ouverte | Helou, Bernard | CCSD

International audience. In this work we propose a generic modeling approach for simulating percutaneous transluminal angioplasty (PTA) endovascular treatment, and evaluating the influence of balloon design, plaque c...

Patient-Specific Finite-Element Simulation of the Insertion of Guidewire During an EVAR Procedure: Guidewire Position Prediction Validation on 28 Cases

Archive ouverte | Gindre, Juliette | CCSD

International audience. Objective: Validation of a numerical method to compute arterial deformations under the insertion of an “extra-siff” guidewire during Endovascular Repair of Abdominal Aortic Aneurysm. Methods:...

Finite element simulation of the insertion of guidewires during an EVAR procedure: example of a complex patient case, a first step toward patient-specific parameterized models

Archive ouverte | Gindre, Juliette | CCSD

International audience. Deformations of the vascular structure due to the insertion of tools during endovascular treatment of aneurysms of the abdominal aorta, unless properly anticipated during the preoperative pla...

Chargement des enrichissements...