Reproducibility of 18F-FDG and 3'-deoxy-3'-18F-fluorothymidine PET tumor volume measurements.. : PET volume determination reproducibility

Archive ouverte

Hatt, Mathieu | Cheze-Le Rest, Catherine | Aboagye, Eric, O. | Kenny, Laura, M. | Rosso, Lula | Turkheimer, Federico, E. | Albarghach, Nidal, M. | Metges, Jean-Philippe | Pradier, Olivier | Visvikis, Dimitris

Edité par CCSD ; Society of Nuclear Medicine -

International audience. The objective of this study was to establish the repeatability and reproducibility limits of several volume-related PET image-derived indices-namely tumor volume (TV), mean standardized uptake value, total glycolytic volume (TGV), and total proliferative volume (TPV)-relative to those of maximum standardized uptake value (SUV(max)), commonly used in clinical practice. METHODS: Fixed and adaptive thresholding, fuzzy C-means, and fuzzy locally adaptive Bayesian methodology were considered for TV delineation. Double-baseline (18)F-FDG (17 lesions, 14 esophageal cancer patients) and 3'-deoxy-3'-(18)F-fluorothymidine ((18)F-FLT) (12 lesions, 9 breast cancer patients) PET scans, acquired at a mean interval of 4 d and before any treatment, were used for reproducibility evaluation. The repeatability of each method was evaluated for the same datasets and compared with manual delineation. RESULTS: A negligible variability of less than 5% was measured for all segmentation approaches in comparison to manual delineation (5%-35%). SUV(max) reproducibility levels were similar to others previously reported, with a mean percentage difference of 1.8% +/- 16.7% and -0.9% +/- 14.9% for the (18)F-FDG and (18)F-FLT lesions, respectively. The best TV, TGV, and TPV reproducibility limits ranged from -21% to 31% and -30% to 37% for (18)F-FDG and (18)F-FLT images, respectively, whereas the worst reproducibility limits ranged from -90% to 73% and -68% to 52%, respectively. CONCLUSION: The reproducibility of estimating TV, mean standardized uptake value, and derived TGV and TPV was found to vary among segmentation algorithms. Some differences between (18)F-FDG and (18)F-FLT scans were observed, mainly because of differences in overall image quality. The smaller reproducibility limits for volume-derived image indices were similar to those for SUV(max), suggesting that the use of appropriate delineation tools should allow the determination of tumor functional volumes in PET images in a repeatable and reproducible fashion.

Suggestions

Du même auteur

Intratumor heterogeneity characterized by textural features on baseline 18F-FDG PET images predicts response to concomitant radiochemotherapy in esophageal cancer.

Archive ouverte | Tixier, Florent | CCSD

International audience. UNLABELLED: (18)F-FDG PET is often used in clinical routine for diagnosis, staging, and response to therapy assessment or prediction. The standardized uptake value (SUV) in the primary or reg...

Prognostic value of 18F-FDG PET image-based parameters in oesophageal cancer and impact of tumour delineation methodology.. Prognostic value of 18F-FDG PET image-based parameters in oesophageal cancer and impact of tumour delineation methodology.: FDG PET indices for survival prediction

Archive ouverte | Hatt, Mathieu | CCSD

International audience. PURPOSE: (18)F-fluorodeoxyglucose (FDG) positron emission tomography (PET) image-derived parameters, such as standardized uptake value (SUV), functional tumour length (TL) and tumour volume (...

PET functional volume delineation: a robustness and repeatability study.. PET functional volume delineation: a robustness and repeatability study.: Robustness of functional volume determination in PET

Archive ouverte | Hatt, Mathieu | CCSD

International audience. PURPOSE: Current state-of-the-art algorithms for functional uptake volume segmentation in PET imaging consist of threshold-based approaches, whose parameters often require specific optimizati...

Chargement des enrichissements...