Trajectory correction enables free-running chemical shift encoded imaging for accurate cardiac proton-density fat fraction quantification at 3T

Archive ouverte

Daudé, Pierre | Troalen, Thomas | Mackowiak, Adèle, L C | Royer, Emilien | Piccini, Davide | Yerly, Jérôme | Pfeuffer, Josef | Kober, Frank | Gouny, Sylviane, Confort | Bernard, Monique | Stuber, Matthias | Bastiaansen, Jessica, a M | Rapacchi, Stanislas

Edité par CCSD ; BioMed Central -

International audience.

Background: Metabolic diseases can negatively alter epicardial fat accumulation and composition, which can be probed using quantitative cardiac chemical shift encoded (CSE) cardiovascular magnetic resonance (CMR) by mapping proton-density fat fraction (PDFF). To obtain motion-resolved high-resolution PDFF maps, we proposed a free-running cardiac CSE-CMR framework at 3T. To employ faster bipolar readout gradients, a correction for gradient imperfections was added using the gradient impulse response function (GIRF) and evaluated on intermediate images and PDFF quantification. Methods: Ten minutes free-running cardiac 3D radial CSE-CMR acquisitions were compared in vitro and in vivo at 3T. Monopolar and bipolar readout gradient schemes provided 8 echoes (TE1/ΔTE = 1.16/1.96 ms) and 13 echoes (TE1/ΔTE = 1.12/1.07 ms), respectively. Bipolar-gradient free-running cardiac fat and water images and PDFF maps were reconstructed with or without GIRF correction. PDFF values were evaluated in silico, in vitro on a fat/water phantom, and in vivo in 10 healthy volunteers and 3 diabetic patients. Results: In monopolar mode, fat-water swaps were demonstrated in silico and confirmed in vitro. Using bipolar readout gradients, PDFF quantification was reliable and accurate with GIRF correction with a mean bias of 0.03% in silico and 0.36% in vitro while it suffered from artifacts without correction, leading to a PDFF bias of

Suggestions

Du même auteur

Deep-Learning Segmentation of Epicardial Adipose Tissue Using Four-Chamber Cardiac Magnetic Resonance Imaging

Archive ouverte | Daudé, Pierre | CCSD

International audience. In magnetic resonance imaging (MRI), epicardial adipose tissue (EAT) overload remains often overlooked due to tedious manual contouring in images. Automated four-chamber EAT area quantificati...

Simultaneous Multi-Slice (SMS) cardiac T1 mapping at 3T using FLASH-MOLLI

Archive ouverte | Rapacchi, Stanislas | CCSD

International audience

Simultaneous multi‐slice cardiac cine with Fourier‐encoded self‐calibration at 7 Tesla

Archive ouverte | Rapacchi, Stanislas | CCSD

International audience. PurposeTo accelerate cardiac cine at 7 tesla using simultaneous multi‐slice (SMS) acquisition with self‐calibration to resolve misalignment between calibration and imaging data due to breathi...

Chargement des enrichissements...