Characterization and optimization of EBT2 radiochromic films dosimetry system for precise measurements of output factors in small fields used in radiotherapy

Archive ouverte

Huet, C. | Dagois, S. | Derreumaux, S. | Trompier, F. | Chenaf, C. | Robbes, I.

Edité par CCSD ; Elsevier -

International audience. The accurate determination of absorbed dose in small photon beams, especially for stereotactic radiation therapy, is a difficult task with commercially available detectors. As these small fields are characterized by high dose gradients, a lack of lateral particle equilibrium and a variation of energy spectra with beam sizes, a dosimeter with high resolution, tissue-equivalence and high precision is required. The new radiochromic film EBT2, which meets these criteria, was fully characterized in Institut de Radioprotection et Sûreté Nucléaire (IRSN) for this application. This type of film was tested with the reading system EPSON Dual Lens Perfection V700 flatbed scanner in transmission mode. Warm-up effects of the scanner were studied as well as the influence of the scanner light. Uniformity of unirradiated and irradiated EBT2 films in terms of pixel value was found to be respectively 0.3% (1 SD) and 0.5% (1 SD). An original, accurate and efficient radiochromic film dosimetry protocol was established. The overall uncertainty for dose measurement with EBT2 films using this protocol was estimated at less than 2% (1 SD). Encouraging measurements of output factors were performed on a Novalis system. © 2011 Elsevier Ltd. All rights reserved.

Consulter en ligne

Suggestions

Du même auteur

Radiation protection of workers associated with secondary neutrons produced by medical linear accelerators

Archive ouverte | Donadille, L. | CCSD

International audience. This paper presents a measurement campaign carried out in six medical electron linear accelerator facilities to evaluate occupational exposure associated with photons and secondary neutrons f...

Dosimetry of the mixed field irradiation facility CALIBAN

Archive ouverte | Trompier, F. | CCSD

International audience. The dosimetry of the experimental reactor CALIBAN was established for photon and neutron components, and the kerma variations were evaluated according to position (distance from the core, hei...

Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones

Archive ouverte | Bassinet, C. | CCSD

Erratum: Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones (Medical Physics (2013) 40 (071725)) ...

Chargement des enrichissements...