Bench Assessment of Expiratory Valve Resistance of Current ICU Ventilators in Dynamic Conditions

Archive ouverte

Pinède, A. | Cour, M. | Degivry, F. | Louis, B. | Argaud, L. | Guérin, C.

Edité par CCSD ; Daedalus Enterprises Inc -

International audience. BACKGROUND: We hypothesized that the lack of benefit of setting a low versus a high PEEP in patients with ARDS may be due in part to differences in the dynamic behavior of the expiratory valve in ventilators. We tested this hypothesis by conducting a bench comparison of the dynamic behavior of expiratory valves on ICU ventilators currently in use. METHODS: We attached 7 ICU ventilators (C5, C6, Carescape, PB980, ServoU, V500, and V680) to the ASL 5000 lung model (passive condition with compliance 20 mL/cm H(2)O and resistance 5 cm H(2)O/L/s) and set in volume controlled mode (tidal volume 0.8 L, breathing frequency 10 breaths/min). Flow and pressure were measured just before the exhalation valve. At PEEP of 5, 10, and 15 cm H(2)O, the median instantaneous expiratory resistance, the time to valve opening, and the pressure time products above or below the values of PEEP (expressed in cm H(2)O × s) were determined. RESULTS: Median instantaneous expiratory resistance values differed between the ventilators and PEEP settings with a significant interaction: at PEEP 5 cm H(2)O, the median (interquartile range) expiratory resistance values were 3.9 (3.5-4.7), 3.0 (3.0-3.1), 20.9 (15.8-24.9), 27.4 (26.5-43.2), 13.8 (13.6-13.9), 4.4 (4.0-4.6), and 34.3 (33.7-33.8) cm H(2)O/L/s, for the C5, C6, Carescape, PB980, ServoU, V500, and V680, respectively. For all the PEEP settings, the corresponding times to valve opening were 0.080 (0.077-0.082), 0.082 (0.080-0.085), 0.110 (0.105-0.110), 0.100 (0.085-1.05), 0.072 (0.062-0.072), 0.145 (0.115-0.150), and 0.075 (0.070-0.080) s, respectively, and pressure-time products were 2.8 (2.1-7.4), 6.8 (6.7-7.3), 2.4 (2.1-2.4), 3.5 (2.7-3.6), 1.8 (1.8-2.1), 2.8 (2.7-2.9), and 5.7 (5.4-5.9) cm H(2)O × s, respectively. CONCLUSIONS: The resistance of active expiratory valves differed significantly between the 7 ICU ventilators tested.

Consulter en ligne

Suggestions

Du même auteur

Simultaneous ventilation in the Covid-19 pandemic. A bench study

Archive ouverte | Guérin, C. | CCSD

International audience. COVID-19 pandemic sets the healthcare system to a shortage of ventilators. We aimed at assessing tidal volume (VT) delivery and air recirculation during expiration when one ventilator is divi...

A Comprehensive Bench Assessment of Automatic Tube Compensation in ICU Ventilators for Better Clinical Management

Archive ouverte | Galerneau, L. M. | CCSD

International audience. BACKGROUND: Automatic tube compensation (ATC) unloads endotracheal tube (ETT) resistance. We conducted a bench assessment of ATC functionality in ICU ventilators to improve clinical managemen...

Regional lung viscoelastic properties in supine and prone position in a porcine model of acute respiratory distress syndrome

Archive ouverte | Guérin, C. | CCSD

International audience. Regional viscoelastic properties of thoracic tissues in acute respiratory distress syndrome (ARDS) and their change with position and positive end-expiratory pressure (PEEP) are unknown. In a...

Chargement des enrichissements...