Deep Learning Denoising Improves and Homogenizes Patient [18F]FDG PET Image Quality in Digital PET/CT

Archive ouverte

Weyts, Kathleen | Quak, Elske | Licaj, Idlir | Ciappuccini, Renaud | Lasnon, Charline | Corroyer-Dulmont, Aurélien | Foucras, Gauthier | Bardet, Stéphane | Jaudet, Cyril

Edité par CCSD ; MDPI -

CERVOXY. International audience. Given the constant pressure to increase patient throughput while respecting radiation protection, global body PET image quality (IQ) is not satisfactory in all patients. We first studied the association between IQ and other variables, in particular body habitus, on a digital PET/CT. Second, to improve and homogenize IQ, we evaluated a deep learning PET denoising solution (Subtle PETTM) using convolutional neural networks. We analysed retrospectively in 113 patients visual IQ (by a 5-point Likert score in two readers) and semi-quantitative IQ (by the coefficient of variation in the liver, CVliv) as well as lesion detection and quantification in native and denoised PET. In native PET, visual and semi-quantitative IQ were lower in patients with larger body habitus (p < 0.0001 for both) and in men vs. women (p ≤ 0.03 for CVliv). After PET denoising, visual IQ scores increased and became more homogeneous between patients (4.8 ± 0.3 in denoised vs. 3.6 ± 0.6 in native PET; p < 0.0001). CVliv were lower in denoised PET than in native PET, 6.9 ± 0.9% vs. 12.2 ± 1.6%; p < 0.0001. The slope calculated by linear regression of CVliv according to weight was significantly lower in denoised than in native PET (p = 0.0002), demonstrating more uniform CVliv. Lesion concordance rate between both PET series was 369/371 (99.5%), with two lesions exclusively detected in native PET. SUVmax and SUVpeak of up to the five most intense native PET lesions per patient were lower in denoised PET (p < 0.001), with an average relative bias of −7.7% and −2.8%, respectively. DL-based PET denoising by Subtle PETTM allowed [18F]FDG PET global image quality to be improved and homogenized, while maintaining satisfactory lesion detection and quantification. DL-based denoising may render body habitus adaptive PET protocols unnecessary, and pave the way for the improvement and homogenization of PET modalities.

Suggestions

Du même auteur

Artificial intelligence-based PET denoising could allow a two-fold reduction in [18F]FDG PET acquisition time in digital PET/CT

Archive ouverte | Weyts, Kathleen | CCSD

CERVOXY. International audience. Purpose We investigated whether artificial intelligence (AI)-based denoising halves PET acquisition time in digital PET/CT. Methods One hundred ninety-five patients referred for [ 18...

Upfront F18-choline PET/CT versus Tc99m-sestaMIBI SPECT/CT guided surgery in primary hyperparathyroidism: the randomized phase III diagnostic trial APACH2

Archive ouverte | Quak, Elske | CCSD

International audience. Background: The common endocrine disorder primary hyperparathyroidism (PHPT) can be cured by surgery. Preoperative localization of parathyroid adenoma (PTA) by imaging is a prerequisite for o...

The Impact of Artificial Intelligence CNN Based Denoising on FDG PET Radiomics

Archive ouverte | Jaudet, Cyril | CCSD

CERVOXY. International audience. Background: With a constantly increasing number of diagnostic images performed each year, Artificial Intelligence (AI) denoising methods offer an opportunity to respond to the growin...

Chargement des enrichissements...