A deep learning model to generate synthetic CT for prostate MR-only radiotherapy dose planning: a multicenter study

Archive ouverte

Tahri, Safaa | Texier, Blanche | Nunes, Jean-Claude | Hemon, Cédric | Lekieffre, Pauline | Collot, Emma | Chourak, Hilda | Le Guevelou, Jennifer | Greer, Peter | Dowling, Jason | Acosta, Oscar | Bessieres, Igor | Marage, Louis | Boue-Rafle, Adrien | de Crevoisier, Renaud | Lafond, Caroline | Barateau, Anais

Edité par CCSD ; Frontiers Media -

International audience. Introduction: For radiotherapy based solely on magnetic resonance imaging (MRI), generating synthetic computed tomography scans (sCT) from MRI is essential for dose calculation. The use of deep learning (DL) methods to generate sCT from MRI has shown encouraging results if the MRI images used for training the deep learning network and the MRI images for sCT generation come from the same MRI device. The objective of this study was to create and evaluate a generic DL model capable of generating sCTs from various MRI devices for prostate radiotherapyMaterials and methods: In total, 90 patients from three centers (30 CT-MR prostate pairs/center) underwent treatment using volumetric modulated arc therapy for prostate cancer (PCa) (60 Gy in 20 fractions). T2 MRI images were acquired in addition to computed tomography (CT) images for treatment planning. The DL model was a 2D supervised conditional generative adversarial network (Pix2Pix). Patient images underwent preprocessing steps, including nonrigid registration. Seven different supervised models were trained, incorporating patients from one, two, or three centers. Each model was trained on 24 CT-MR prostate pairs. A generic model was trained using patients from all three centers. To compare sCT and CT, the mean absolute error in Hounsfield units was calculated for the entire pelvis, prostate, bladder, rectum, and bones. For dose analysis, mean dose differences of D99% for CTV, V95% for PTV, Dmax for rectum and bladder, and 3D gamma analysis (local, 1%/1 mm) were calculated from CT and sCT. Furthermore, Wilcoxon tests were performed to compare the image and dose results obtained with the generic model to those with the other trained models.Results: Considering the image results for the entire pelvis, when the data used for the test comes from the same center as the data used for training, the results were not significantly different from the generic model. Absolute dose differences were less than 1 Gy for the CTV D99% for every trained model and center. The gamma analysis results showed nonsignificant differences between the generic and monocentric models.Conclusion: The accuracy of sCT, in terms of image and dose, is equivalent to whether MRI images are generated using the generic model or the monocentric model. The generic model, using only eight MRI-CT pairs per center, offers robust sCT generation, facilitating PCa MRI-only radiotherapy for routine clinical use.

Suggestions

Du même auteur

Computed tomography synthesis from magnetic resonance imaging using cycle Generative Adversarial Networks with multicenter learning

Archive ouverte | Texier, Blanche | CCSD

International audience. Background and Purpose: Addressing the need for accurate dose calculation in MRI-only radiotherapy, the generation of synthetic Computed Tomography (sCT) from MRI has emerged. Deep learning (...

3D Unsupervised deep learning method for magnetic resonance imaging-to-computed tomography synthesis in prostate radiotherapy

Archive ouverte | Texier, Blanche | CCSD

International audience. Background and purpose: Magnetic resonance imaging (MRI)-to-computed tomography (CT) synthesis is essential in MRI-only radiotherapy workflows, particularly through deep learning techniques k...

Quality assurance for MRI-only radiation therapy: A voxel-wise population-based methodology for image and dose assessment of synthetic CT generation methods

Archive ouverte | Chourak, Hilda | CCSD

International audience. The quality assurance of synthetic CT (sCT) is crucial for safe clinical transfer to an MRI-only radiotherapy planning workflow. The aim of this work is to propose a population-based process ...

Chargement des enrichissements...