Remotely detuned receiver coil for high-resolution interventional cardiac magnetic resonance imaging

Archive ouverte

Marhabaie, Sina | Delcey, Marylene | El Hamrani, Dounia | Vaillant, Fanny | Ginefri, Jean-Christophe | Ozenne, Valery | Abell, Emma | Poirier-Quinot, Marie | Quesson, Bruno

Edité par CCSD ; Frontiers Media -

International audience. IntroductionInterventional cardiac MRI in the context of the treatment of cardiac arrhythmia requires submillimeter image resolution to precisely characterize the cardiac substrate and guide the catheter-based ablation procedure in real-time. Conventional MRI receiver coils positioned on the thorax provide insufficient signal-to-noise ratio (SNR) and spatial selectivity to satisfy these constraints.MethodsA small circular MRI receiver coil was developed and evaluated under different experimental conditions, including high-resolution MRI anatomical and thermometric imaging at 1.5 T. From the perspective of developing a therapeutic MR-compatible catheter equipped with a receiver coil, we also propose alternative remote active detuning techniques of the receiver coil using one or two cables. Theoretical details are presented, as well as simulations and experimental validation.ResultsAnatomical images of the left ventricle at 170 µm in-plane resolution are provided on ex vivo beating heart from swine using a 2 cm circular receiver coil. Taking advantage of the increase of SNR at its vicinity (up to 35 fold compared to conventional receiver coils), real-time MR-temperature imaging can reach an uncertainty below 0.1°C at the submillimetric spatial resolution. Remote active detuning using two cables has similar decoupling efficiency to conventional on-site decoupling, at the cost of an acceptable decrease in the resulting SNR.DiscussionThis study shows the potential of small dimension surface coils for minimally invasive therapy of cardiac arrhythmia intraoperatively guided by MRI. The proposed remote decoupling approaches may simplify the construction process and reduce the cost of such single-use devices.

Suggestions

Du même auteur

Remotely detuned receiver coil for high-resolution interventional cardiac magnetic resonance imaging

Archive ouverte | Marhabaie, Sina | CCSD

International audience. Introduction Interventional cardiac MRI in the context of the treatment of cardiac arrhythmia requires submillimeter image resolution to precisely characterize the cardiac substrate and guide...

Characterization of the Septal Discontinuity in Ex-Vivo Human Hearts Using Diffusion Tensor Imaging: The Potential Structural Determinism Played by Fiber Orientation in Clinical Phenotype of Laminopathy Patients

Archive ouverte | Cabanis, Pierre | CCSD

International audience. A key component of the cardiac conduction system is the atrioventricular (AV) node where the His bundle (HB) divides into the right (RBB) and left (LBB) bundles branches. The RBB is connected...

Assessment of Cardiac Toxicity of Manganese Chloride for Cardiovascular Magnetic Resonance

Archive ouverte | Lamonzie, Elodie | CCSD

International audience. MRI is widely used in cardiology to characterize the structure and function of the heart. Currently, gadolinium-based contrast agents are widely used to improve sensitivity and specificity of...

Chargement des enrichissements...