CT dose optimization for the detection of pulmonary arteriovenous malformation (PAVM): A phantom study

Archive ouverte

Greffier, J. | Boccalini, S. | Beregi, J.P. | Vlassenbroek, A. | Vuillod, A. | Dupuis-Girod, S. | Boussel, L. | Douek, P. | Si-Mohamed, Salim

Edité par CCSD ; Elsevier -

International audience. To determine the lowest suitable dose level for the detection of pulmonary arteriovenous malformation (PAVM) using a task-based image quality assessment.Material and methods: A phantom was scanned using the standard chest protocol (STD) and 4 other ultra-low dose protocols (ULD) using various kVp. Raw data were reconstructed using level 5 of the hybrid iterative reconstruction algorithm (iDose4) for the STD protocol, and level 6 of iDose4 and levels 1 to 3 of model-based iterative reconstruction (IMR) for the ULD protocols. Both quantitative criteria and qualitative analysis were used to compare protocols. Noise-power-spectrum and Task-based transfer function were computed using imQuest software. The detectability-index (d') was computed for the detection of PAVM. A subjective analysis was performed by 2 chest radiologists to validate the image-quality obtained on the anthropomorphic phantom for all protocols.Results: Similar d' values were found for ULD-140 using iDose4 6 compared to STD protocol. Greater d' values were found for all ULD protocols using IMR compared to STD. Subjective image quality was rated as acceptable to excellent for ULD-140 and ULD-120 for all reconstruction types, for ULD-100 and ULD-80 using IMR2, and for ULD-100 using IMR1. Image smoothing was poor for IMR3 for ULD-100 and ULD-80. Finally, the ULD-80 protocol reconstructed with IMR2 was chosen for the detection of PAVM. With this protocol, the dose (CTDIvol of 0.3mGy) was reduced by 91% compared with the STD protocol.Conclusion: A dose level as low as 0.3mGy reconstructed with IMR2 provides an image quality suitable for the detection of PAVM.

Suggestions

Du même auteur

Head-to-head comparison of lung perfusion with dual-energy CT and SPECT-CT

Archive ouverte | Si-Mohamed, S. | CCSD

International audience

Impact of iterative reconstructions on image quality and detectability of focal liver lesions in low-energy monochromatic images

Archive ouverte | Greffier, J. | CCSD

International audience. Purpose: To assess the impact of iterative reconstructions on image quality and detectability of focal liver lesions in low-energy monochromatic images from a Fast kV-Switching Dual Energy CT...

Assessment of peak skin dose in interventional cardiology: A comparison between Gafchromic film and dosimetric software em.dose

Archive ouverte | Greffier, J. | CCSD

International audience. OBJECTIVE:To compare the use of a dose mapping software to Gafchromic film measurement for a simplified peak skin dose (PSD) estimation in interventional cardiology procedure.METHODS:The stud...

Chargement des enrichissements...