0 avis
Manifestation of aberrations in full-field optical coherence tomography
Archive ouverte
Edité par CCSD ; Optical Society of America - OSA Publishing -
23 pages, 10 figures. International audience. We report on a theoretical model for image formation in full-field optical coherence tomography (FFOCT). Because the spatial incoherence of the illumination acts as a virtual confocal pinhole in FFOCT, its imaging performance is equivalent to a scanning time-gated coherent confocal microscope. In agreement with optical experiments enabling a precise control of aberrations, FFOCT is shown to have nearly twice the resolution of standard imaging at moderate aberration level. Beyond a rigorous study on the sensitivity of FFOCT with respect to aberrations, this theoretical model paves the way towards an optimized design of adaptive optics schemes for high-resolution and deep imaging of biological tissues.