Development of a protocol for small beam bi-dimensional dose distribution measurements with radiochromic films

Archive ouverte

Moignier, C. | Huet, C. | Barraux, V. | Fontaine, Julien | Loiseau, C. | Sebe-Mercier, K. | Batalla, A. | Makovicka, L.

Edité par CCSD ; Elsevier -

International audience. Gafchromic™ films have become popular due to their ease of use and their near water equivalence. This last property is crucial for stereotactic small beam dosimetry as demonstrated in recent papers. An accurate bi-dimensional dose measurement with Gafchromic™ films is very challenging mainly because of the non-uniformity response of flatbed scanners (used for films digitalization) and their own non-uniformity. The first proposal of this work is to develop bi-dimensional protocol for small beams and evaluate the associated uncertainty. The second proposal is to validate this protocol for the bi-dimensional measurements of treatment plans performed with the CyberKnife® system. First, the uniformity of an Epson V700 flatbed scanner and a batch of EBT3 Gafchromic™ films has been investigated. A "four films" dosimeter was designed to reduce the errors (statistic and systematic) due to their non-uniformity. Then, the "four films" dosimeter protocol in both a homogeneous (RW3 material) and heterogeneous (RW3, lung-like and bone-like materials) phantoms has been used to measure the bi-dimensional dose distributions of three simple CyberKnife® treatment plans. Two tumor locations (middle of the lung and near lung/bone interface) were considered for the heterogeneous phantom. These plans were achieved with the 10 mm fixed collimator and modeled with the PENELOPE Monte Carlo code in order to calculate accurate dose distributions. Finally, the "four films" bi-dimensional dose distributions were compared to the PENELOPE Monte Carlo simulations. Regarding the uncertainty associated to the bi-dimensional dose measurement protocol, the relative standard deviation σD on the dose was 1.2% in the range from 0.5 to 4.0 Gy. Regarding the protocol validation on CyberKnife® treatment plans, a very good agreement was found with all measurement points passing the {3% - 3 mm} Gamma Index criteria. © 2016 Elsevier Ltd. All rights reserved.

Consulter en ligne

Suggestions

Du même auteur

Study of commercial detector responses in non-equilibrium small photon fields of a 1000 MU/min CyberKnife system

Archive ouverte | Huet, C. | CCSD

International audience. Purpose The purpose of this study was to analyze the detector responses in non-equilibrium small photon fields. Methods Five detectors (PTW 31014 ionization chamber, PTW 60016, PTW 60017 and ...

Determination of the kfclin,fmsr Qclin,Qmsr correction factors for detectors used with an 800 MU/min CyberKnife® system equipped with fixed collimators and a study of detector response to small photon beams using a Monte Carlo method

Archive ouverte | Moignier, C. | CCSD

International audience. Purpose In a previous work, output ratio (ORdet) measurements were performed for the 800 MU/min CyberKnife® at the Oscar Lambret Center (COL, France) using several commercially available dete...

Characterization of the gafchromic EBT3 films for dose distribution measurements in stereotactic radiotherapy

Archive ouverte | Huet, C. | CCSD

International audience. Radiochromic film dosimetry is a promising technique, but at this time there are some artefacts, including non-uniformity, energy dependence and scanner artefacts. Accurate dosimetry with rad...

Chargement des enrichissements...