Study of commercial detector responses in non-equilibrium small photon fields of a 1000 MU/min CyberKnife system

Archive ouverte

Huet, C. | Moignier, C. | Barraux, V. | Loiseau, C. | Sebe-Mercier, K. | Batalla, A. | Clairand, I.

Edité par CCSD ; Elsevier -

International audience. Purpose The purpose of this study was to analyze the detector responses in non-equilibrium small photon fields. Methods Five detectors (PTW 31014 ionization chamber, PTW 60016, PTW 60017 and Sun Nuclear EDGE diodes and PTW 60003 diamond detector) and one passive dosimeter (Harshaw micro-LiF) as well as a 1000 MU/min CyberKnife were modeled with the PENELOPE Monte Carlo code. Field factors, ΩQclin,Qmsrfclin,fmsr were calculated and perturbations due to volume averaging effect, active material effect and coating effect were quantified for the five detectors and passive dosimeter. Results The PTW 31014 ionization chamber under-response is mainly due to the fluence perturbation caused by the presence of air as detecting material. Regarding diodes, the high density materials used in their active volume and in their coating is responsible for their over-response. Regarding the PTW 60003 diamond, its under-response for the 5 mm field size is due to a large volume averaging effect whereas for largest field sizes a nearly perfect compensation between the volume averaging effect and the material effect due to the diamond density occurs. Despite its small size, a volume averaging effect was observed for the micro-LiF for the 5 mm field size. Conclusion Perturbations due to volume averaging effect, active material effect and coating effect were investigated and quantified for five active detectors. Since these perturbations can cause opposite effects, wrong conclusions may be drawn regarding the radiological water-equivalence of detectors. Thus, we recommend performing such a study for each novel detector available on the market. © 2016 Associazione Italiana di Fisica Medica.

Consulter en ligne

Suggestions

Du même auteur

Development of a protocol for small beam bi-dimensional dose distribution measurements with radiochromic films

Archive ouverte | Moignier, C. | CCSD

International audience. Gafchromic™ films have become popular due to their ease of use and their near water equivalence. This last property is crucial for stereotactic small beam dosimetry as demonstrated in recent ...

Characterization of the gafchromic EBT3 films for dose distribution measurements in stereotactic radiotherapy

Archive ouverte | Huet, C. | CCSD

International audience. Radiochromic film dosimetry is a promising technique, but at this time there are some artefacts, including non-uniformity, energy dependence and scanner artefacts. Accurate dosimetry with rad...

Digital applicator by 3D printing in contact brachytherapy. Applicateur numérique par impression tridimensionnelle en curiethérapie de contact

Archive ouverte | Lecornu, M. | CCSD

International audience. Brachytherapy of skin tumours uses custom applicators that are manufactured manually. The integration of 3D printing customization of applicators during hidh dose rate brachytherapy planning ...

Chargement des enrichissements...