Evaluation of the Needle Positioning Accuracy of a Light Puncture Robot Under MRI Guidance: Results of a Clinical Trial on Healthy Volunteers

Archive ouverte

Ghelfi, Julien | Moreau-Gaudry, Alexandre | Hungr, Nikolai | Fouard, Céline | Veron, Baptiste | Medici, Maud | Chipon, Emilie | Cinquin, Philippe | Bricault, Ivan

Edité par CCSD ; Cardiovascular and Interventional Radiological Society of Europe (CIRSE). -

International audience. PURPOSE:To assess the accuracy of Light Puncture Robot (LPR) as a patient-mounted robot, in positioning a sham needle under MRI guidance for abdominal percutaneous interventions.MATERIALS AND METHODS:This monocentric, prospective and non-controlled study was approved by the ethics review board. The study evaluated the accuracy of LPR V3 to achieve a virtual puncture in 20 healthy volunteers. Three trajectories were tried on each volunteer, under 3-T MRI guidance.RESULTS:Accuracy under 5 mm in attaining a 10 cm-deep target was reached in 72% of attempts after 2 robot motions with a median error of 4.1 mm [2.1; 5.1]. Median procedure time for one trajectory was 12.9 min [10.2; 18.0] and median installation time was 9.0 min [6.0; 13.0].CONCLUSION:LPR accuracy in the deployment of a sham needle inside the MRI tunnel and its setup time are promising. Further studies need to be conducted to confirm these results before clinical trials.

Consulter en ligne

Suggestions

Du même auteur

Computer assisted electromagnetic navigation improves accuracy in computed tomography guided interventions: A prospective randomized clinical trial

Archive ouverte | Durand, Pierre | CCSD

International audience. PURPOSE:To assess the accuracy and usability of an electromagnetic navigation system designed to assist Computed Tomography (CT) guided interventions.MATERIALS AND METHODS:120 patients requir...

Design and Validation of a CT- and MRI-Guided Robot for Percutaneous Needle Procedures

Archive ouverte | Hungr, Nikolai | CCSD

International audience. This paper describes the design and evaluation of the second generation of a robotic system called the light puncture robot for thoracic and abdominopelvic interventional radiology procedures...

First Clinical Experience in Urologic Surgery with a Novel Robotic Lightweight Laparoscope Holder.

Archive ouverte | Long, Jean-Alexandre | CCSD

International audience. Purpose: To report the feasibility and the safety of a surgeon-controlled robotic endoscope holder in laparoscopic surgery. Materials and methods: From March 2010 to September 2010, 20 patien...

Chargement des enrichissements...