Multicenter outpatient dinner/overnight reduction of hypoglycemia and increased time of glucose in target with a wearable artificial pancreas using modular model predictive control in adults with type 1 diabetes

Archive ouverte

del Favero, S. | Place, Jérôme | Kropff, J. | Messori, M. | Keith-Hynes, P. | Visentin, R. | Monaro, M. | Galasso, S. | Boscari, F. | Toffanin, C. | Di Palma, F. | Lanzola, G. | Scarpellini, S. | Farret, Anne | Kovatchev, B. | Avogaro, A. | Bruttomesso, D. | Magni, L. | Devries, J.H. | Cobelli, C. | Renard, Eric | Consortium, Ap@home

Edité par CCSD ; Wiley -

International audience. AIMS:To test in an outpatient setting the safety and efficacy of continuous subcutaneous insulin infusion (CSII) driven by a modular model predictive control (MMPC) algorithm informed by continuous glucose monitoring (CGM) measurement.METHODS:13 patients affected by type 1 diabetes participated to a non-randomized outpatient 42-h experiment that included two evening meals and overnight periods (in short, dinner & night periods). CSII was patient-driven during dinner & night period 1 and MMPC-driven during dinner&night period 2. The study was conducted in hotels, where patients could move around freely. A CGM system (G4 Platinum; Dexcom Inc., San Diego, CA, USA) and insulin pump (AccuChek Combo; Roche Diagnostics, Mannheim, Germany) were connected wirelessly to a smartphone-based platform (DiAs, Diabetes Assistant; University of Virginia, Charlottesville, VA, USA) during both periods.RESULTS:A significantly lower percentage of time spent with glucose levels <3.9 mmol/l was achieved in period 2 compared with period 1: 1.96 ± 4.56% vs 12.76 ± 15.84% (mean ± standard deviation, p < 0.01), together with a greater percentage of time spent in the 3.9-10 mmol/l target range: 83.56 ± 14.02% vs 62.43 ± 29.03% (p = 0.04). In addition, restricting the analysis to the overnight phases, a lower percentage of time spent with glucose levels <3.9 mmol/l (1.92 ± 4.89% vs 12.7 ± 19.75%; p = 0.03) was combined with a greater percentage of time spent in 3.9-10 mmol/l target range in period 2 compared with period 1 (92.16 ± 8.03% vs 63.97 ± 2.73%; p = 0.01). Average glucose levels were similar during both periods.CONCLUSIONS:The results suggest that MMPC managed by a wearable system is safe and effective during evening meal and overnight. Its sustained use during this period is currently being tested in an ongoing randomized 2-month study.

Consulter en ligne

Suggestions

Du même auteur

Safety of outpatient closed-loop control: first randomized crossover trials of a wearable artificial pancreas

Archive ouverte | Kovatchev, B. P. | CCSD

International audience. OBJECTIVE We estimate the effect size of hypoglycemia risk reduction on closed-loop control (CLC) versus open-loop (OL) sensor-augmented insulin pump therapy in supervised outpatient setting....

Psychological outcomes of evening and night closed-loop insulin delivery under free living conditions in people with Type 1 diabetes: a 2-month randomized crossover trial

Archive ouverte | Kropff, J. | CCSD

International audience. AIM:To assess the impact on fear of hypoglycaemia and treatment satisfaction with an artificial pancreas system used for 2 consecutive months, as well as participant acceptance of the artific...

Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions

Archive ouverte | Kropff, J. | CCSD

International audience. AIMS:To assess the accuracy and reliability of the two most widely used continuous glucose monitoring (CGM) systems.METHODS:We studied the Dexcom®G4 Platinum (DG4P; Dexcom, San Diego, CA, USA...

Chargement des enrichissements...