An Automated Sample Preparation Instrument to Accelerate Positive Blood Cultures Microbial Identification by MALDI-TOF Mass Spectrometry (Vitek®MS)

Archive ouverte

Broyer, Patrick | Perrot, Nadine | Rostaing, Hervé | Blaze, Jérome | Pinston, Frederic | Gervasi, Gaspard | Charles, Marie-Hélène | Dachaud, Fabien | Dachaud, Jacques | Moulin, Frederic | Cordier, Sylvain | Dauwalder, Olivier | Meugnier, Hélène | Vandenesch, Francois

Edité par CCSD ; Frontiers Media -

International audience. Sepsis is the leading cause of death among patients in intensive care units (ICUs) requiring an early diagnosis to introduce efficient therapeutic intervention. Rapid identification (ID) of a causative pathogen is key to guide directed antimicrobial selection and was recently shown to reduce hospitalization length in ICUs. Direct processing of positive blood cultures by MALDI-TOF MS technology is one of the several currently available tools used to generate rapid microbial ID. However, all recently published protocols are still manual and time consuming, requiring dedicated technician availability and specific strategies for batch processing. We present here a new prototype instrument for automated preparation of Vitek\textregisteredMS slides directly from positive blood culture broth based on an "all-in-one" extraction strip. This bench top instrument was evaluated on 111 and 22 organisms processed using artificially inoculated blood culture bottles in the BacT/ALERT\textregistered 3D (SA/SN blood culture bottles) or the BacT/ALERT VirtuoTM system (FA/FN Plus bottles), respectively. Overall, this new preparation station provided reliable and accurate Vitek MS species-level identification of 87% (Gram-negative bacteria = 85%, Gram-positive bacteria = 88%, and yeast = 100%) when used with BacT/ALERT\textregistered 3D and of 84% (Gram-negative bacteria = 86%, Gram-positive bacteria = 86%, and yeast = 75%) with Virtuo\textregistered instruments, respectively. The prototype was then evaluated in a clinical microbiology laboratory on 102 clinical blood culture bottles and compared to routine laboratory ID procedures. Overall, the correlation of ID on monomicrobial bottles was 83% (Gram-negative bacteria = 89%, Gram-positive bacteria = 79%, and yeast = 78%), demonstrating roughly equivalent performance between manual and automatized extraction methods. This prototype instrument exhibited a high level of performance regardless of bottle type or BacT/ALERT system. Furthermore, blood culture workflow could potentially be improved by converting direct ID of positive blood cultures from a batch-based to real-time and "on-demand" process.

Suggestions

Du même auteur

Rapid bacterial identification, resistance, virulence and type profiling using selected reaction monitoring mass spectrometry

Archive ouverte | Charretier, Yannick | CCSD

International audience. Mass spectrometry (MS) in Selected Reaction Monitoring (SRM) mode is proposed for in-depth characterisation of microorganisms in a multiplexed analysis. Within 60-80 minutes, the SRM method p...

Legionella pneumophila sequence type 1/Paris pulsotype subtyping by spoligotyping.

Archive ouverte | Ginevra, Christophe | CCSD

International audience. Endemic strains of Legionella pneumophila sequence type 1 (ST1), in particular the ST1/Paris pulsotype, are dispersed worldwide and represent about 10% of culture-proven clinical cases of Leg...

Species identification of staphylococci by amplification and sequencing of the tuf gene compared to the gap gene and by matrix-assisted laser desorption ionization time-of-flight mass spectrometry.

Archive ouverte | Bergeron, Marie | CCSD

Staphylococcal species, notably, coagulase-negative staphylococci (CoNS), are frequently misidentified using phenotypic methods. The partial nucleotide sequences of the tuf and gap genes were determined in 47 reference strains to ...

Chargement des enrichissements...