Assessment of peak skin dose in interventional cardiology: A comparison between Gafchromic film and dosimetric software em.dose

Archive ouverte

Greffier, J. | van Ngoc Ty, C. | Bonniaud, G. | Moliner, G. | Ledermann, B. | Schmutz, L. | Cornillet, L. | Cayla, G. | Beregi, J.P. | Pereira, F.

Edité par CCSD ; Elsevier -

International audience. OBJECTIVE:To compare the use of a dose mapping software to Gafchromic film measurement for a simplified peak skin dose (PSD) estimation in interventional cardiology procedure.METHODS:The study was conducted on a total of 40 cardiac procedures (20 complex coronary angioplasty of chronic total occlusion (CTO) and 20 coronary angiography and coronary angioplasty (CA-PTCA)) conducted between January 2014 to December 2015. PSD measurement (PSDFilm) was obtained by placing XR-RV3 Gafchromic under the patient's back for each procedure. PSD (PSDem.dose) was computed with the software em.dose©. The calculation was performed on the dose metrics collected from the private dose report of each procedure. Two calculation methods (method A: fluoroscopic kerma equally spread on cine acquisition and B: fluoroscopic kerma is added to one air Kerma cine acquisition that contributes to the PSD) were used to calculate the fluoroscopic dose contribution as fluoroscopic data were not recorded in our interventional room. Statistical analyses were carried out to compare PSDFilm and PSDem.dose.RESULTS:The PSDFilm median (1st quartile; 3rd quartile) was 0.251(0.190;0.336)Gy for CA-PTCA and 1.453(0.767;2.011)Gy for CTO. For method-A, the PSDem.dose was 0.248(0.182;0.369)Gy for CA-PTCA and 1.601(0.892;2.178)Gy for CTO, and 0.267(0.223;0.446)Gy and 1.75 (0.912;2.584)Gy for method-B, respectively. For the two methods, the correlation between PSDFilm and PSDem.dose was strong. For all cardiology procedures investigated, the mean deviation between PSDFilm and PSDem.dose was 3.4±21.1% for method-A and 17.3%±23.9% for method-B.CONCLUSION:The dose mapping software is convenient to calculate peak skin dose in interventional cardiology.

Consulter en ligne

Suggestions

Du même auteur

Assessment of Patient's Peak Skin Dose Using Gafchromic Films During Interventional Cardiology Procedures: Routine Experience Feedback

Archive ouverte | Greffier, J. | CCSD

International audience. To assess the interest of Gafchromic films in detection of patient's peak skin dose (PSD) in interventional cardiology. A prospective study of 112 patients was conducted (July-December 2015)....

Impact of the calibration conditions of XR-RV3 films on peak skin dose measurements in interventional radiology

Archive ouverte | Greffier, J. | CCSD

International audience. In this study, the impact of radiochromic films (XR-RV3) calibration on PSD measurements was investigated under various peak kilovoltage (kVp) and additional filtration conditions. Films were...

[What is the place of bioresorbable vascular scaffolds in setting of ST-segment elevation myocardial infarction?].. Quelle place pour les stents biorésorbables dans la prise en charge du syndrome coronarien aigu avec sus-décalage du segment ST ?

Archive ouverte | Lattuca, B. | CCSD

International audience. A 50-year-old woman was admitted for an inferior ST-segment elevation myocardial infarction; immediate coronary angiogram revealed a subocclusive stenosis of the right coronary artery. After ...

Chargement des enrichissements...