Impact of the calibration conditions of XR-RV3 films on peak skin dose measurements in interventional radiology

Archive ouverte

Greffier, J. | van Ngoc Ty, C. | Agelou, Mathieu | Coulot, J. | Ledermann, B. | Ovtchinnikoff, S. | Pereira, F. | Beregi, J. | Bonniaud, G.

Edité par CCSD ; Oxford University Press (OUP) -

International audience. In this study, the impact of radiochromic films (XR-RV3) calibration on PSD measurements was investigated under various peak kilovoltage (kVp) and additional filtration conditions. Films were calibrated free-in-air for six beam qualities with Allura Xper FD20 system (Philips). Six calibration curves (CCs) were constructed. Each beam quality was characterized in terms of mean energy (ME) in the air, with table, with table and water phantom using Monte Carlo simulations. A cohort of 155 patient films from cardiology (37) and vascular (118) procedures were read with each CC. Routine calibration beam quality was taken as reference (DoseNorm). Overall, it was observed that for a wider ME difference between the exposed film and the CC used, a larger deviation (from -28% to +41%) was observed. The choice of beam quality for the calibration is a key point when additional filtration and kVp are automatically controlled in clinical conditions.

Suggestions

Du même auteur

Assessment of peak skin dose in interventional cardiology: A comparison between Gafchromic film and dosimetric software em.dose

Archive ouverte | Greffier, J. | CCSD

International audience. OBJECTIVE:To compare the use of a dose mapping software to Gafchromic film measurement for a simplified peak skin dose (PSD) estimation in interventional cardiology procedure.METHODS:The stud...

Assessment of patient's peak skin dose during abdominopelvic embolization using radiochromic (Gafchromic) films

Archive ouverte | Larbi, A. | CCSD

International audience. PURPOSE: To assess the value of the routine use of radiochromic films in abdominopelvic embolization procedures to improve patient follow-up. METHODS: A total of 55~patients who underwent tra...

Assessment of Patient's Peak Skin Dose Using Gafchromic Films During Interventional Cardiology Procedures: Routine Experience Feedback

Archive ouverte | Greffier, J. | CCSD

International audience. To assess the interest of Gafchromic films in detection of patient's peak skin dose (PSD) in interventional cardiology. A prospective study of 112 patients was conducted (July-December 2015)....

Chargement des enrichissements...