Real-time Volumetric MRI Thermometry of Focused Ultrasound Ablation in vivo : A feasibility study in pig liver and kidney

Archive ouverte

Quesson, Bruno | Christophe, Laurent | Maclair, G. | Denis de Senneville, Baudouin | Mougenot, Charles | Ries, Mario | Carteret, Thibaut | Rullier, A. | Moonen, Chrit, T. W.

Edité par CCSD ; Wiley -

International audience. MR thermometry offers the possibility to precisely guide high-intensity focused ultrasound (HIFU) for the noninvasive treatment of kidney and liver tumours. The objectives of this study were to demonstrate therapy guidance by motion-compensated, rapid and volumetric MR temperature monitoring and to evaluate the feasibility of MR-guided HIFU ablation in these organs. Fourteen HIFU sonications were performed in the kidney and liver of five pigs under general anaesthesia using an MR-compatible Philips HIFU platform prototype. HIFU sonication power and duration were varied. Volumetric MR thermometry was performed continuously at 1.5 T using the proton resonance frequency shift method employing a multi-slice, single-shot, echo-planar imaging sequence with an update frequency of 2.5 Hz. Motion-related suceptibility artefacts were compensated for using multi-baseline reference images acquired prior to sonication. At the end of the experiment, the animals were sacrificed for macroscopic and microscopic examinations of the kidney, liver and skin. The standard deviation of the temperature measured prior to heating in the sonicated area was approximately 1-C in kidney and liver, and 2.5-C near the skin. The maximum temperature rise was 30-C for a sonication of 1.2 MHz in the liver over 15 s at 300 W. The thermal dose reached the lethal threshold (240CEM 43) in two of six cases in the kidney and four of eight cases in the liver, but remained below this value in skin regions in the beam path. These findings were in agreement with histological analysis. Volumetric thermometry allows real-time monitoring of the temperature at the target location in liver and kidney, as well as in surrounding tissues. Thermal ablation was more difficult to achieve in renal than in hepatic tissue even using higher acoustic energy, probably because of a more efficient heat evacuation in the kidney by perfusion.

Suggestions

Du même auteur

MR-thermometry for monitoring tumor ablation

Archive ouverte | Denis de Senneville, Baudouin | CCSD

International audience. Local thermal therapies are increasingly used in the clinic for tissue ablation. During energy deposition, the actual tissue temperature is difficult to estimate since physiological processes...

A method for MRI guidance of intercostal high intensity focused ultrasound ablation in the liver

Archive ouverte | Quesson, Bruno | CCSD

International audience. High Intensity Focused Ultrasound (HIFU) seems promising for the non invasive treatment of liver tumors. However, the presence of ribs in the HIFU beam path remains problematic since they can...

Magnetic Resonance Temperature Imaging

Archive ouverte | Denis de Senneville, Baudouin | CCSD

International audience. Continuous, real-time 3D temperature mapping during a hyperthermic procedure may provide enhanced safety by visualizing temperature maps in and around the treated region, improved efficiency ...

Chargement des enrichissements...