Damage to inner ear structure during cochlear implantation: Correlation between insertion force and radio-histological findings in temporal bone specimens

Archive ouverte

de Seta, Daniele | Torres, Renato | Russo, Francesca | Ferrary, Evelyne | Kazmitcheff, Guillaume | Heymann, Dominique | Amiaud, Jérôme | Sterkers, Olivier | Bernardeschi, Daniele | Nguyen, Yann

Edité par CCSD ; Elsevier -

International audience. Cochlear implant insertion should be as least traumatic as possible in order to reduce trauma to the cochlear sensory structures. The force applied to the cochlea during array insertion should be controlled to limit insertion-related damage. The relationship between insertion force and histological traumatism remains to be demonstrated. Twelve freshly frozen cadaveric temporal bones were implanted with a long straight electrodes array through an anterior extended round window insertion using a motorized insertion tool with real-time measurement of the insertion force. Anatomical parameters, measured on a pre-implantation cone beam CT scan, position of the array and force metrics were correlated with post-implantation scanning electron microscopy images and histological damage assessment. An atraumatic insertion occurred in six cochleae, a translocation in five cochleae and a basilar membrane rupture in one cochlea. The translocation always occurred in the 150-to 180-degree region. In the case of traumatic insertion, different force profiles were observed with a more irregular curve arising from the presence of an early peak force (30 ± 18.2 mN). This corresponded approximately to the first point of contact of the array with the lateral wall of the cochlea. Atraumatic and traumatic insertions had significantly different force values at the same depth of insertion (p < 0.001, two-way ANOVA), and significantly different regression lines (y ¼ 1.34x þ 0.7 for atraumatic and y ¼ 3.37x þ 0.84 for traumatic insertion, p < 0.001, ANCOVA). In the present study, the insertion force was correlated with the intracochlear trauma. The 150-to 180-degree region represented the area at risk for scalar translocation for this straight electrodes array. Insertion force curves with different sets of values were identified for traumatic and atraumatic insertions; these values should be considered during motorized insertion of an implant so as to be able to modify the insertion parameters (e.g axis of insertion) and facilitate preservation of endocochlear structures.

Suggestions

Du même auteur

Quality of insertion in cochlear implants : a clinical and temporal bone study. Qualité de l'insertion des implants cochléaires : une étude clinique et sur os temporal

Archive ouverte | de Seta, Daniele | CCSD

The cochlear implantation represents the standard technique to restore the hearing in totallydeafened persons, and the indications during the last years are widening also to patients withresidual hearing or presenting single sided...

Influence of electrode array stiffness and diameter on hearing in cochlear implanted guinea pig

Archive ouverte | Drouillard, Mylène | CCSD

International audience. During cochlear implantation, electrode array translocation and trauma should be avoided to preserve residual hearing. The aim of our study was to evaluate the effect of physical parameters o...

Meniett device in Meniere disease: Randomized, double-blind, placebo-controlled multicenter trial

Archive ouverte | Russo, Francesca Yoshie | CCSD

International audience. Objective: To evaluate the efficacy of portable Meniett low-pressure pulse generator (Medtronic Xomed, Jacksonville, FL) in Meniere disease.Study Design: Randomized, double-blind, placebo-con...