Intra-operative ultrasound hand-held strain imaging for the visualization of ablations produced in the liver with a toroidal HIFU transducer: first in vivo results.

Archive ouverte

Chenot, Jérémy | Melodelima, David | N'Djin, William Apoutou | Souchon, Rémi | Rivoire, Michel | Chapelon, Jean-Yves

Edité par CCSD ; IOP Publishing -

International audience. The use of hand-held ultrasound strain imaging for the intra-operative real-time visualization of HIFU (high-intensity focused ultrasound) ablations produced in the liver by a toroidal transducer was investigated. A linear 12 MHz ultrasound imaging probe was used to obtain radiofrequency signals. Using a fast cross-correlation algorithm, strain images were calculated and displayed at 60 frames s(-1), allowing the use of hand-held strain imaging intra-operatively. Fourteen HIFU lesions were produced in four pigs. Intra-operative strain imaging of HIFU ablations in the liver was feasible owing to the high frame rate. The correlation between dimensions measured on gross pathology and dimensions measured on B-mode images and on strain images were R = 0.72 and R = 0.94 respectively. The contrast between ablated and non-ablated tissue was significantly higher (p < 0.05) in the strain images (22 dB) than in the B-mode images (9 dB). Strain images allowed equivalent or improved definition of ablated regions when compared with B-mode images. Real-time intra-operative hand-held strain imaging seems to be a promising complement to conventional B-mode imaging for the guidance of HIFU ablations produced in the liver during an open procedure. These results support that hand-held strain imaging outperforms conventional B-mode ultrasound and could potentially be used for the assessment of thermal therapies.

Suggestions

Du même auteur

Using a toroid transducer for thermal ablation by high-intensity-focused ultrasound increases the coagulated volume.

Archive ouverte | Melodelima, David | CCSD

International audience. Here, we report that a new design of High Intensity Focused Ultrasound transducer can significantly enlarge the coagulated volume over short periods of time, and that treatment in the liver c...

A New High Channels Density Ultrasound Platform for Advanced 4D Cardiac Imaging

Archive ouverte | Petrusca, Lorena | CCSD

International audience. Background, Motivation and ObjectiveA novel ultrasound (US) platform with high channels density that offers flexibility, precision and open access is a pre-requisite to open new frontiers in ...

A Tumor-mimic Model for Evaluating the Accuracy of HIFU Preclinical Studies: An In Vivo Study.

Archive ouverte | N'Djin, William Apoutou | CCSD

International audience. To date, the efficacy of ablative technologies such as HIFU for the treatment of liver tumors in humans has been studied in animal models without tumors or in small animals like rats and rabb...

Chargement des enrichissements...