64-Elements intraluminal ultrasound cylindrical phased array for transesophageal thermal ablation under fast MR temperature mapping: An ex vivo study

Archive ouverte

Melodelima, David | Salomir, Rares | Mougenot, Charles | Moonen, Chrit | Cathignol, Dominique

Edité par CCSD ; American Association of Physicists in Medicine -

This work was undertaken to investigate the feasibility of using a cylindrical phased array for transoesophaeal thermal ablation under Magnetic Resonance (MR) imaging guidance. Sixty four transducers (0.45-mm wide by 15-mm tall), operating at 4.6 MHz, were spread around the periphery of a 10.6-mm diameter cylinder. The head of the applicator was covered with a 65­µm thick latex balloon attached using watertight seals. This envelope was inflated with degassed water to provide acoustic coupling between the transducer and the tissues. The underlying operating principle of this applicator is to rotate a plane ultrasound beam electronically. For this purpose, eight adjacent transducers were excited with appropriate delay times so as to generate a plane wave. The exposure direction was changed by exciting a different set of eight elements. Ex vivo experiments conducted on 47 samples of pig liver under MR temperature monitoring demonstrated the ability of this applicator to generate cylindrical or sector-based coagulation necroses at depths up to 19 mm with excellent angular precision by applying 20 W/cm2. MR-thermometry was performed in “real-time” with segmented echo-planar imaging gradient echo sequences. The temporal resolution was approximately 3s / image. The average value for the temperature baseline in liver tissue close to the applicator was 0.3°C (±0.6°C). The thermal dose delivered in tissues was computed on-line during temperature imaging. Excellent MR compatibility was demonstrated, all MR acquisitions were performed without susceptibility artefacts or radiofrequency interferences with the ultrasound device. Thermal lesions identified on post-treatment follow up showed good correlation with on line MR thermometry data. The individual differences between measurements performed visually and using MRI thermal dose maps were about 11% of volume. This study demonstrated the feasibility of thermal ablation using a phased array intraluminal ultrasound applicator and on line MR monitoring.

Suggestions

Du même auteur

Real-time adaptive methods for treatment of mobile organs by MRI-controlled high-intensity focused ultrasound

Archive ouverte | Denis de Senneville, Baudouin | CCSD

International audience. Focused ultrasound is a unique and non-invasive technique for the local deposition of thermal energy deep inside the body. MRI guidance offers the additional benefits of excellent target visu...

A method for MRI guidance of intercostal high intensity focused ultrasound ablation in the liver

Archive ouverte | Quesson, Bruno | CCSD

International audience. High Intensity Focused Ultrasound (HIFU) seems promising for the non invasive treatment of liver tumors. However, the presence of ribs in the HIFU beam path remains problematic since they can...

Interstitial devices for minimally invasive thermal ablation by high-intensity ultrasound.

Archive ouverte | Lafon, Cyril | CCSD

10 pages. Interstitial ultrasound applicators have been proposed for treating deep-seated tumours that cannot be reached with extra-corporeal high-intensity focused ultrasound. In addition, interstitial ultrasound o...

Chargement des enrichissements...