Frequency Drift in MR Spectroscopy at 3T.

Archive ouverte

Hui, S. C. N. | Mikkelsen, M. | Zöllner, H. J. | Ahluwalia, V. | Alcauter, S. | Baltusis, L. | Barany, D. A. | Barlow, L. R. | Becker, R. | Berman, J. I. | Berrington, A. | Bhattacharyya, P. K. | Blicher, J. U. | Bogner, W. | Brown, M. S. | Calhoun, V. D. | Castillo, R. | Cecil, K. M. | Choi, Y. B. | Chu, W. C. W. | Clarke, W. T. | Craven, A. R. | Cuypers, K. | Dacko, M. | de la Fuente-Sandoval, C. | Desmond, P. | Domagalik, A. | Dumont, Julien | Duncan, N. W. | Dydak, U. | Dyke, K. | Edmondson, D. A. | Ende, G. | Ersland, L. | Evans, C. J. | Fermin, A. S. R. | Ferretti, A. | Fillmer, A. | Gong, T. | Greenhouse, I. | Grist, J. T. | Gu, M. | Harris, A. D. | Hat, K. | Heba, S. | Heckova, E. | Hegarty, J. P. | Heise, K. F. | Jacobson, A. | Jansen, J. F. A. | Jenkins, C. W. | Johnston, S. J. | Juchem, C. | Kangarlu, A. | Kerr, A. B. | Landheer, K. | Lange, T. | Lee, P. | Levendovszky, S. R. | Limperopoulos, C. | Liu, F. | Lloyd, W. | Lythgoe, D. J. | Machizawa, M. G. | Macmillan, E. L. | Maddock, R. J. | Manzhurtsev, A. V. | Martinez-Gudino, M. L. | Miller, J. J. | Mirzakhanian, H. | Moreno-Ortega, M. | Mullins, P. G. | Near, J. | Noeske, R. | Nordhøy, W. | Oeltzschner, G. | Osorio-Duran, R. | Otaduy, M. C. G. | Pasaye, E. H. | Peeters, R. | Peltier, S. J. | Pilatus, U. | Polomac, N. | Porges, E. C. | Pradhan, S. | Prisciandaro, J. J. | Puts, N. A. | Rae, C. D. | Reyes-Madrigal, F. | Roberts, T. P. L. | Robertson, C. E. | Rosenberg, J. T. | Rotaru, D. G. | O'Gorman Tuura, R. L. | Saleh, M. G. | Sandberg, K. | Sangill, R. | Schembri, K. | Schrantee, A. | Semenova, N. A. | Singel, D. | Sitnikov, R. | Smith, J. | Song, Y. | Stark, C. | Stoffers, D. | Swinnen, S. P. | Tain, R. | Tanase, C. | Tapper, S. | Tegenthoff, M. | Thiel, T. | Thioux, M. | Truong, P. | van Dijk, P. | Vella, N. | Vidyasagar, R. | Vovk, A. | Wang, G. | Westlye, L. T. | Wilbur, T. K. | Willoughby, W. R. | Wilson, M. | Wittsack, H. J. | Woods, A. J. | Wu, Y. C. | Xu, J. | Lopez, M. Y. | Yeung, D. K. W. | Zhao, Q. | Zhou, X. | Zupan, G. | Edden, R. A. E.

Edité par CCSD ; Elsevier -

International audience. PurposeHeating of gradient coils and passive shim components is a common cause of instability in the B0 field, especially when gradient intensive sequences are used. The aim of the study was to set a benchmark for typical drift encountered during MR spectroscopy (MRS) to assess the need for real-time field-frequency locking on MRI scanners by comparing field drift data from a large number of sites.MethodA standardized protocol was developed for 80 participating sites using 99 3T MR scanners from 3 major vendors. Phantom water signals were acquired before and after an EPI sequence. The protocol consisted of: minimal preparatory imaging; a short pre-fMRI PRESS; a ten-minute fMRI acquisition; and a long post-fMRI PRESS acquisition. Both pre- and post-fMRI PRESS were non-water suppressed. Real-time frequency stabilization/adjustment was switched off when appropriate. Sixty scanners repeated the protocol for a second dataset. In addition, a three-hour post-fMRI MRS acquisition was performed at one site to observe change of gradient temperature and drift rate. Spectral analysis was performed using MATLAB. Frequency drift in pre-fMRI PRESS data were compared with the first 5:20 minutes and the full 30:00 minutes of data after fMRI. Median (interquartile range) drifts were measured and showed in violin plot. Paired t-tests were performed to compare frequency drift pre- and post-fMRI. A simulated in vivo spectrum was generated using FID-A to visualize the effect of the observed frequency drifts. The simulated spectrum was convolved with the frequency trace for the most extreme cases. Impacts of frequency drifts on NAA and GABA were also simulated as a function of linear drift. Data from the repeated protocol were compared with the corresponding first dataset using Pearson's and intraclass correlation coefficients (ICC).ResultsOf the data collected from 99 scanners, 4 were excluded due to various reasons. Thus, data from 95 scanners were ultimately analyzed. For the first 5:20 min (64 transients), median (interquartile range) drift was 0.44 (1.29) Hz before fMRI and 0.83 (1.29) Hz after. This increased to 3.15 (4.02) Hz for the full 30 min (360 transients) run. Average drift rates were 0.29 Hz/min before fMRI and 0.43 Hz/min after. Paired t-tests indicated that drift increased after fMRI, as expected (p < 0.05). Simulated spectra convolved with the frequency drift showed that the intensity of the NAA singlet was reduced by up to 26%, 44 % and 18% for GE, Philips and Siemens scanners after fMRI, respectively. ICCs indicated good agreement between datasets acquired on separate days. The single site long acquisition showed drift rate was reduced to 0.03 Hz/min approximately three hours after fMRI.DiscussionThis study analyzed frequency drift data from 95 3T MRI scanners. Median levels of drift were relatively low (5-min average under 1 Hz), but the most extreme cases suffered from higher levels of drift. The extent of drift varied across scanners which both linear and nonlinear drifts were observed.

Suggestions

Du même auteur

Genetic architecture of subcortical brain structures in 38,851 individuals

Archive ouverte | Satizabal, C. L. | CCSD

International audience. Subcortical brain structures are integral to motion, consciousness, emotions and learning. We identified common genetic variation related to the volumes of the nucleus accumbens, amygdala, br...

Contemporary Outcomes After Partial Resection Of Infected Aortic Grafts

Archive ouverte | Janko, M. | CCSD

International audience. IntroductionAortic graft infection remains a considerable clinical challenge, and it is unclear which variables are associated with adverse outcomes among patients undergoing partial resectio...

Zinc incorporation in the miliolid foraminifer Pseudotriloculina rotunda under laboratory conditions

Archive ouverte | Nardelli, Maria Pia | CCSD

International audience. The incorporation rate of Zn into the calcareous tests of Pseudotriloculina rotunda was investigated in culture in order to evaluate the possibility of using Zn/Ca ratios as a pollution proxy...

Chargement des enrichissements...