A performance evaluation of sthemO 301 coagulation analyzer and associated reagents

Archive ouverte

Pontis, Adeline | Delanoe, Mélanie | Schilliger, Nathalie | Carlo, Audrey | Gueret, Dominique Pierre | Nédelec-Gac, Fabienne | Gouin-Thibault, Isabelle

Edité par CCSD ; Wiley-Blackwell -

International audience. AIM: The study objective was to evaluate the performance of sthemO 301 system and to compare it with the analyzer used in our university hospital laboratory (STA R Max® 2), for a selection of hemostasis parameters. METHODS: Method comparison (according to CLSI EP09-A3), carryover (according to CLSI H57-A), APTT sensitivity to heparin (according to CLSI H47-A2), HIL level assessment, and productivity were performed using leftover samples from our laboratory (n > 1000). Commercial quality control materials were used to evaluate precision (according to CLSI EP15-A3) and accuracy. The assays tested on sthemO 301 were: PT, APTT (silica and kaolin activators), fibrinogen (Fib), thrombin time (TT), chromogenic and clotting protein C (PC) activity, and von Willebrand factor antigen (VWF:Ag) levels. RESULTS: All intra-assay and inter-assay precision CVs were below the maximal precision limit proposed by the French Group for Hemostasis and Thrombosis (GFHT). Accuracy was verified with bias below GFHT criteria and most Z-scores were between -2 and +2. No clinically relevant carryover was detected. Silica APTT reagent sensitivity to unfractionated heparin was moderate, as expected. Productivity results were consistent over the 10 repeats performed. The overall agreement between the two systems was excellent for all assays, with Spearman rank correlation coefficient all above 0.9 and slopes of Passing-Bablok correlation near 1 and intercepts close to 0. CONCLUSION: For the methods tested, sthemO 301 system met all the criteria to implement a novel coagulation analyzer in the laboratory and result comparability with STA R Max® 2 was good.

Consulter en ligne

Suggestions

Du même auteur

Impact of age on in vitro metabolism of clopidogrel: a potential explanation for high on-treatment platelet reactivity in the elderly?

Archive ouverte | Pontis, Adeline | CCSD

International audience. BACKGROUND: High on-treatment platelet reactivity has been reported in 30% of patients on clopidogrel and 50% in elderly patients; however, little is known about the mechanisms of this biolog...

Tinzaparin, an alternative to subcutaneous unfractionated heparin, in patients with severe and end-stage renal impairment: a retrospective observational single-center study

Archive ouverte | Gouin-Thibault, Isabelle | CCSD

International audience. Background: Tinzaparin could be easier to manage than unfractionated heparin in patients with severe renal impairment. However, clinical and pharmacologic data regarding its use in such patie...

Apixaban and rivaroxaban in obese patients treated for venous thromboembolism: Drug levels and clinical outcomes

Archive ouverte | Ballerie, Alice | CCSD

International audience. BACKGROUND: Direct oral anticoagulants (DOAC) use remains challenging in obese patients treated for Venous-Thrombo-Embolism (VTE) due to the paucity of prospective and dedicated studies. OBJE...