Optimization of image quality and accuracy of low iodine concentration quantification as function of kVp pairs for abdominal imaging using dual-source CT: A phantom study

Archive ouverte

Dabli, Djamel | Frandon, Julien | Hamard, Aymeric | Belaouni, Asmaa | Addala, Takieddine | Beregi, Jean-Paul | Greffier, Joël

Edité par CCSD ; Elsevier -

International audience. Purpose: To determine the suitable kVp pair for optimal image quality of the virtual monochromatic images (VMIs) and iodine quantification accuracy at low concentration, using a third generation dual-source CT (DSCT).Materials and methods: Multi-energy CT phantoms with and without body rings were scanned with a DSCT using four kVp pairs (tube "A"/"B" voltage): 100/Sn150, 90/Sn150, 80/Sn150 and 70/Sn150 kVp. The reference mAs was adjusted to obtain a CTDIvol close to 11 mGy. HU values accuracy (RMSDHU), noise (SD) and contrast-to-noise ratio (CNR) of iodine inserts of 0.5, 1, 2 and 5 mg/mL concentrations were assessed on VMIs at 40/50/60/70 keV. Iodine quantification accuracy was assessed using the RMSDiodine and iodine bias (IBiodine).Results: The RMSDHU decreased when the tube "A" voltage increased. The mean noise value increased significantly with tube "A" voltage (p < 0.001) but decreased between 80/Sn150 and 90/Sn150 kVp for the small phantom (1.1 ± 0.1%; p = 0.047). The CNR significantly decreased with tube "A" voltage (p < 0.001), except between 80/Sn150 and 90/Sn150 kVp for all inserts and between 90/Sn150 kVp and 100/Sn150 kVp for the 1.0 and 0.5 mg/mL inserts in the large phantom. In the small phantom, no significant difference was found between 80/Sn150 kVp and 90/Sn150 kVp for all inserts and between 80/Sn150, 90/Sn150 and 100/Sn150 kVp for the 1 and 0.5 mg/mL inserts. The RMSDiodine and IBiodine decreased as the tube "A" voltage of the kVp pair increased.Conclusion:The kVp pair of 70/Sn150 led to better image quality in VMIs and sufficient iodine accuracy.

Suggestions

Du même auteur

Optimization of image quality and accuracy of low iodine concentration quantification as function of dose level and reconstruction algorithm for abdominal imaging using dual-source CT: A phantom study

Archive ouverte | Dabli, Djamel | CCSD

International audience. Purpose: The purpose of this study was to assess the impact of advanced modeled iterative reconstruction (ADMIRE) algorithm and dose levels on the accuracy of Hounsfield unit (HU) measurement...

Effect of a new deep learning image reconstruction algorithm for abdominal computed tomography imaging on image quality and dose reduction compared with two iterative reconstruction algorithms: a phantom study

Archive ouverte | Greffier, Joël | CCSD

International audience. Background: New reconstruction algorithms based on deep learning have been developed to correct the image texture changes related to the use of iterative reconstruction algorithms. The purpos...

Comparison of two deep learning image reconstruction algorithms in chest CT images: A task-based image quality assessment on phantom data

Archive ouverte | Greffier, Joël | CCSD

International audience. Purpose: The purpose of this study was to compare the effect of two deep learning image reconstruction (DLR) algorithms in chest computed tomography (CT) with different clinical indications.M...

Chargement des enrichissements...