Comparison of Holmium:YAG and Thulium Fiber Lasers on the Risk of Laser Fiber Fracture

Archive ouverte

Uzan, Audrey | Chiron, Paul | Panthier, Frédéric | Haddad, Mattieu | Berthe, Laurent | Traxer, Olivier | Doizi, Steeve

Edité par CCSD ; MDPI -

International audience. Objectives: To compare the risk of laser fiber fracture between Ho:YAG laser and Thulium Fiber Laser (TFL) with different laser fiber diameters, laser settings, and fiber bending radii.Methods: Lengths of 200, 272, and 365 μm single use fibers were used with a 30 W Ho:YAG laser and a 50 W Super Pulsed TFL. Laser fibers of 150 µm length were also tested with the TFL only. Five different increasingly smaller bend radii were tested: 1, 0.9, 0.75, 0.6, and 0.45 cm. A total of 13 different laser settings were tested for the Ho:YAG laser: six fragmentation settings with a short pulse duration, and seven dusting settings with a long pulse duration. A total of 33 different laser settings were tested for the TFL. Three laser settings were common two both lasers: 0.5 J × 12 Hz, 0.8 J × 8 Hz, 2 J × 3 Hz. The laser was activated for 5 min or until fiber fracture. Each measurement was performed ten times.Results: While fiber failures occurred with all fiber diameters with Ho:YAG laser, none were reported with TFL. Identified risk factors of fiber fracture with the Ho:YAG laser were short pulse and high energy for the 365 µm fibers (p = 0.041), but not for the 200 and 272 µm fibers (p = 1 and p = 0.43, respectively). High frequency was not a risk factor of fiber fracture. Fiber diameter also seemed to be a risk factor of fracture. The 200 µm fibers broke more frequently than the 272 and 365 µm ones (p = 0.039). There was a trend for a higher number of fractures with the 365 µm fibers compared to the 272 µm ones, these occurring at a larger bend radius, but this difference was not significant.Conclusion: TFL appears to be a safer laser regarding the risk of fiber fracture than Ho:YAG when used with fibers in a deflected position.

Suggestions

Du même auteur

Evaluating Noise Emissions of Endourological Lasers: A Comparative Analysis of Ho:YAG, Tm:YAG, and Thulium Fiber Laser Systems

Archive ouverte | Moretto, Stefano | CCSD

International audience. Objective:To evaluate the noise levels of Holmium:YAG (Ho:YAG), Thulium Fiber (TFL), and pulsed Thulium:YAG (p-Tm:YAG) lasers across various settings, focusing on compliance with safety thres...

Evaluation of a free 3D software for kidney stones’ surgical planning: “kidney stone calculator” a pilot study

Archive ouverte | Panthier, Frédéric | CCSD

International audience. Introduction: Kidney Stone Calculator (KSC) is a free, three-dimensional (3D) planning software for flexible ureteroscopy(fURS) with Holmium:YAG(Ho:YAG) endocorporeal lithotripsy (EL). KSC pr...

Laser Fiber Displacement Velocity during Tm-Fiber and Ho:YAG Laser Lithotripsy: Introducing the Concept of Optimal Displacement Velocity

Archive ouverte | Panthier, Frederic | CCSD

International audience. Background: Endocorporeal laser lithotripsy (EL) during flexible ureteroscopy (URS-f) often uses “dusting” settings with “painting” technique. The displacement velocity of the laser fiber (LF...

Chargement des enrichissements...