Monitoring of cardiovascular physiology augmented by a patient-specific biomechanical model during general anesthesia. A proof of concept study

Archive ouverte

Le Gall, Arthur | Vallée, Fabrice | Pushparajah, Kuberan | Hussain, Tarique | Mebazaa, Alexandre | Chapelle, Dominique | Gayat, Etienne | Chabiniok, Radomir

Edité par CCSD ; Public Library of Science -

International audience. During general anesthesia (GA), direct analysis of arterial pressure or aortic flow waveforms may be inconclusive in complex situations. Patient-specific biomechanical models, based on data obtained during GA and capable to perform fast simulations of cardiac cycles, have the potential to augment hemodynamic monitoring. Such models allow to simulate Pressure-Volume (PV) loops and estimate functional indicators of cardiovascular (CV) system, e.g. ventricular-arterial coupling (Vva), cardiac efficiency (CE) or myocardial contractility, evolving throughout GA. In this prospective observational study, we created patient-specific biomechanical models of heart and vasculature of a reduced geometric complexity for n=45 patients undergoing GA, while using transthoracic echocardiography and aortic pressure and flow signals acquired in the beginning of GA (baseline condition). If intraoperative hypotension (IOH) appeared, diluted norepinephrine (NOR) was administered and the model readjusted according to the measured aortic pressure and flow signals. Such patients were a posteriori assigned into a so-called hypotensive group. The accuracy of simulated mean aortic pressure (MAP) and stroke volume (SV) at baseline were in accordance with the guidelines for the validation of new devices or reference measurement methods in all patients. After NOR administration in the hypotensive group, the percentage of concordance with 10% exclusion zone between measurement and simulation was > 95% for both MAP and SV. The modeling results showed a decreased Vva (0.64±0.37 vs 0.88±0.43; p=0.039) and an increased CE (0.8±0.1 vs 0.73±0.11; p=0.042) in hypotensive vs normotensive patients. Furthermore, Vva increased by 92±101%, CE decreased by 13±11% (p < 0.001 for both) and contractility increased by 14±11% (p = 0.002) in the hypotensive group post-NOR administration. In this work we demonstrated the application of fast-running patient-specific biophysical models to estimate PV loops and functional indicators of CV system using clinical data available during GA. The work paves the way for model-augmented hemodynamic monitoring at operating theatres or intensive care units to enhance the information on patient-specific physiology.

Suggestions

Du même auteur

Coupling data measured in operating theatre with patient-specific biomechanical model of heart and vessels to augment haemodynamic monitoring of patients undergoing general anaesthesia. Exploitation conjointe d'un modèle biomécanique patient-spécifique de coeur et vaisseaux et de données mesurées au bloc opératoire pour augmenter le monitorage hémodynamique des patients sous anesthésie générale

Archive ouverte | Le Gall, Arthur | CCSD

Haemodynamic evaluation of anaesthetised patients in operating theatre or in intensive care units can be challenging. Indeed, monitoring devices can be inaccurate and/or invasive. Physicians have to evaluate the usefulness of the ...

A biomechanics-based parametrized cardiac end-diastolic pressure–volume relationship for accurate patient-specific calibration and estimation

Archive ouverte | Chapelle, Dominique | CCSD

International audience. A simple power law has been proposed in the pioneering work of Klotz et al. (Am J Physiol Heart Circ Physiol 291(1):H403–H412, 2006) to approximate the end-diastolic pressure–volume relations...

Minimally-invasive estimation of patient-specific end-systolic elastance using a biomechanical heart model

Archive ouverte | Le Gall, Arthur | CCSD

International audience. The end-systolic elastance $(Ees)$-the slope of the end-systolic pressure-volume relationship (ESPVR) at the end of ejection phase-has become a reliable indicator of myocardial functional sta...