Radiation doses to patients in interventional coronary procedures-estimation of updated national reference levels by dose audit

Archive ouverte

Georges, J.-L. | Belle, L. | Etard, C. | Azowa, J.-B. | Albert, F. | Pansieri, M. | Monsegu, J. | Barbou, F. | Trouillet, C. | Leddet, P. | Livarek, B. | Marcaggi, X. | Hanssen, M. | Cattan, S.

Edité par CCSD ; Oxford University Press (OUP) -

International audience. The objective of this study was to estimate the French national updated reference levels (RLs) for coronary angiography (CA) and percutaneous coronary intervention (PCI) by a dose audit from a large data set of unselected procedures and in standard sized patients. Kerma-area product (PKA), air kerma at interventional point (Ka,r), fluoroscopy time (FT), and the number of registered frames (NFs) and runs (NRs) were collected from 51 229 CAs and 42 222 PCIs performed over a 12-month period at 61 French hospitals. RLs estimated by the 75th percentile in CAs and PCIs performed in unselected patients were 36 and 78Gy.cm2 for PKA, 498 and 1285 mGy for Ka,r, 6 and 15 min for FT, and 566 and 960 for NF, respectively. These values were consistent with the RLs calculated in standard-sized patients. The large difference in dose between sexes leads us to pro pose specific RLs in males and females. The results suggest a trend for a time-course reduction in RLs for interventional cor onary procedures.

Consulter en ligne

Suggestions

Du même auteur

SESAME A TOOL FOR NUMERICAL DOSIMETRIC RECONSTRUCTION OF PATIENTS OVEREXPOSURES IN INTERVENTIONAL RADIOLOGY

Archive ouverte | Huet, C. | CCSD

International audience. Radiation overexposure accidents are rare but can have severe health consequences. Evaluating the dose received by the patient is a crucial step in the medical management. For that purpose, f...

Establishment of diagnostic reference levels in cardiac ct in france A need for patient dose optimisation

Archive ouverte | Mafalanka, F. | CCSD

International audience. The objective of this study was to propose diagnostic reference levels (DRLs) for coronary computed tomography angiography (CCTA), in the context of a large variability in patient radiation d...

[Atypical MRI image of Tako-Tsubo syndrome: a case report]

Archive ouverte | Melay, M. | CCSD

International audience

Chargement des enrichissements...