How to solve the problem of co-elution between two compounds in liquid chromatography through the first UV derivative spectrum. A trial on alternative plasticizers to di(2-ethylhexyl) phthalate.

Archive ouverte

Masse, Morgane | Genay, Stéphanie | Feutry, Frédéric | Simon, Nicolas | Barthélémy, Christine | Sautou, Valérie | Décaudin, Bertrand | Odou, Pascal | Group, For The Armed Study

Edité par CCSD ; Elsevier -

International audience. To meet new regulations, alternative plasticizers to di(2-ethylhexyl) phthalate (DEHP) are now commonly used in the manufacturing of medical devices. These are: acetyl tri-n-butyl citrate (ATBC), bis (2-ethylhexyl)adipate (DEHA), dioctyl terephtalate (DEHT), di-isononylphtalate (DINP), diisononylcyclohexane-1.2-dicarboxylate (DINCH) and trioctyltrimellilate (TOTM). An HPLC-UV analysis was previously developed to characterize four of them. However, two compounds were systematically co-eluated: DEHP with DEHA and DEHT with DINP. The first derivative of UV spectra and photodiode array detection allow the quantification of DEHA and DINP. Moreover, for each plasticizer, maximum wavelength absorbance was chosen to be as specific as possible. Quantification ranged from 0.3 to 750 µg/mL according to the plasticizer. The assays were validated by analysis of variance. Our method was validated by determining the following parameters: specificity, linearity, limits of detection and quantification. The relative biases were inferior to 5% for ATBC, DEHP, DEHA and DINCH and inferior to 10% for DEHT, DINP and TOTM. Plasticizers were extracted with tetrahydrofuran and methanol. The developed method was then used to determine the composition of plasticizers in several medical devices used in clinical service. The major plasticizers were quantified from 19% to 40% w/w, traces of DEHT were found in six medical devices and DEHP in five.

Consulter en ligne

Suggestions

Du même auteur

Identification and quantification by 1H nuclear magnetic resonance spectroscopy of seven plasticizers in PVC medical devices.

Archive ouverte | Genay, Stéphanie | CCSD

International audience. Medical devices are generally made of polyvinyl chloride plasticized by six authorized plasticizers as alternatives to di-(2-ethylhexyl)-phthalate (DEHP) classified as reprotoxic class 1b. Th...

Long-term stability of ready-to-use 1-mg/mL midazolam solution.

Archive ouverte | Gilliot, Sixtine | CCSD

International audience. PurposeMidazolam is a benzodiazepine derivative commonly used in intensive care units to control sedation. Its use requires dilution of a 5-mg/mL commercial solution to a target concentration...

To what extent do the storage conditions of polyether‐based polyurethane have an impact on diazepam delivery?. To what extent do the storage conditions of polyether-based polyurethane have an impact on diazepam delivery?

Archive ouverte | Maiguy‐foinard, Aurélie | CCSD

International audience. Interactions between medical device material and the drug itself have been evoked for polyurethane and may lead to underdosing. Polyurethane, sterilization mode, and the crosslinking level of...