Are the Current Doppler Echocardiography Criteria Able to Discriminate Mitral Bileaflet Mechanical Heart Valve Malfunction? An In Vitro Study

Archive ouverte

Evin, Morgane | Guivier-Curien, Carine | Pibarot, Philippe | Kadem, Lyes | Rieu, Régis

Edité par CCSD ; Wiley -

International audience. Malfunction of bileaflet mechanical heart valves in the mitral position could either be due to patient-prosthesis mismatch (PPM) or leaflet obstruction. The aim of this article is to investigate the validity of current echocardiographic criteria used for diagnosis of mitral prosthesis malfunction, namely maximum velocity, mean transvalvular pressure gradient, effective orifice area, and Doppler velocity index. In vitro testing was performed on a double activation left heart duplicator. Both PPM and leaflet obstruction were investigated on a St. Jude Medical Master. PPM was studied by varying the St. Jude prosthesis size (21, 25, and 29 mm) and stroke volume (70 and 90 mL). Prosthesis leaflet obstruction was studied by partially or totally blocking the movement of one valve leaflet. Mitral flow conditions were altered in terms of E/A ratios (0.5, 1.0, and 1.5) to simulate physiologic panel of diastolic function. Maximum velocity, effective orifice area, and Doppler velocity index are shown to be insufficient to distinguish normal from malfunctioning St. Jude prostheses. Doppler velocity index and effective orifice area were 1.3 +/- 0.49 and 1.83 +/- 0.43 cm(2) for testing conditions with no malfunction below the 2.2 and 2 cm(2) thresholds (1.19 cm(2) for severe PPM and 1.23 cm(2) for fully blocked leaflet). The mean pressure gradient reached 5 mm Hg thresholds for several conditions of severe PPM only (6.9 mm Hg and mean maximum velocity value: 183.4 cm/s) whereas such value was never attained in the case of leaflet obstruction. In the case of leaflet obstruction, the maximum velocity averaged over the nine pulsed-wave Doppler locations increased by 38% for partial leaflet obstruction and 75% for a fully blocked leaflet when compared with normal conditions. Current echocardiographic criteria might be suboptimal for the detection of bileaflet mechanical heart valve malfunction. Further developments and investigations are required in order to further improve current guidelines.

Consulter en ligne

Suggestions

Du même auteur

Localized Transvalvular Pressure Gradients in Mitral Bileaflet Mechanical Heart Valves and Impact on Gradient Overestimation by Doppler

Archive ouverte | Morgane, Evin | CCSD

International audience. BackgroundIt has been reported that localized high velocity may be recorded by continuous-wave Doppler interrogation through the smaller central orifices of bileaflet mechanical heart valves ...

Mitral valve-in-valve hemodynamic performance: An in vitro study

Archive ouverte | Evin, Morgane | CCSD

International audience. Objectives: The valve-in-valve (VinV) procedure may be used in high-risk patients with failed mitral surgical bioprostheses. The objective of this in vitro study was to assess the hemodynamic...

Effects of Hemodynamic Conditions and Valve Sizing on Leaflet Bending Stress in Self‐Expanding Transcatheter Aortic Valve: An In‐vitro Study

Archive ouverte | Stanova, Viktória | CCSD

International audience

Chargement des enrichissements...