Optimization of intra-voxel incoherent motion imaging at 3.0 Tesla for fast liver examination.

Archive ouverte

Leporq, B. | Saint-Jalmes, Hervé | Rabrait, C. | Pilleul, F. | Guillaud, O. | Dumortier, J. | Scoazec, J.-Y. | Beuf, O.

Edité par CCSD ; Wiley-Blackwell -

International audience. BACKGROUND: Optimization of multi b-values MR protocol for fast intra-voxel incoherent motion imaging of the liver at 3.0 Tesla. METHODS: A comparison of four different acquisition protocols were carried out based on estimated IVIM (DSlow , DFast , and f) and ADC-parameters in 25 healthy volunteers. The effects of respiratory gating compared with free breathing acquisition then diffusion gradient scheme (simultaneous or sequential) and finally use of weighted averaging for different b-values were assessed. An optimization study based on Cramer-Rao lower bound theory was then performed to minimize the number of b-values required for a suitable quantification. The duration-optimized protocol was evaluated on 12 patients with chronic liver diseases RESULTS: No significant differences of IVIM parameters were observed between the assessed protocols. Only four b-values (0, 12, 82, and 1310 s.mm(-2) ) were found mandatory to perform a suitable quantification of IVIM parameters. DSlow and DFast significantly decreased between nonadvanced and advanced fibrosis (P < 0.05 and P < 0.01) whereas perfusion fraction and ADC variations were not found to be significant. CONCLUSION: Results showed that IVIM could be performed in free breathing, with a weighted-averaging procedure, a simultaneous diffusion gradient scheme and only four optimized b-values (0, 10, 80, and 800) reducing scan duration by a factor of nine compared with a nonoptimized protocol. Preliminary results have shown that parameters such as DSlow and DFast based on optimized IVIM protocol can be relevant biomarkers to distinguish between nonadvanced and advanced fibrosis. J. Magn. Reson. Imaging 2014. © 2014 Wiley Periodicals, Inc.

Consulter en ligne

Suggestions

Du même auteur

Combinaison d'une méthode de quantification de la graisse intra-hépatique avec l'imagerie du mouvement incohérent intra-voxel 3.OT pour la distinction entre la stéatose pure et la stéatose hépatite

Archive ouverte | Leporq, B. | CCSD

National audience

Characterization of a dedicated double loop, endoluminal coil for anal sphincter MR imaging at 1.5 T and 3 T

Archive ouverte | Verret, J.-M. | CCSD

International audience. MRI has proven its usefulness in the prediction of surgical anterior anal repair that cannot be done with the reference endosonographic exam. Conventional endorectal coils are often based on ...

Noninvasive diagnosis and prognosis of liver cirrhosis: a comparison of biological scores elastometry and metabolic liver function tests

Archive ouverte | Forestier, J. | CCSD

Chargement des enrichissements...