Validation of an automated blood culture system for sterility testing of cell therapy products.

Archive ouverte

Hocquet, Didier | Sauget, Marlène | Roussel, Sandrine | Malugani, Caroline | Pouthier, Fabienne | Morel, Pascal | Gbaguidi-Haore, Houssein | Bertrand, Xavier | Grenouillet, Frédéric

Edité par CCSD ; Elsevier -

International audience. BACKGROUND AIMS: Automated blood culture systems are widely used for the detection of microorganisms in cell therapy products. However, they are not validated by the manufacturers for this purpose. The aim of this study was to assess the ability of the Bactec system (Becton-Dickinson, Le Pont-De-Claix, France) to detect the microorganisms that could contaminate cell therapy products. METHODS: Three types of vials and conditions were tested: Plus Aerobic/F and Anaerobic/F media incubated at 35°C and Mycosis IC/F medium incubated at 30°C. All vials were incubated 10 days. We tested 18 microorganisms, including slow growers and some with fastidious nutritional requirements (10 bacteria, four yeasts, four filamentous fungi), each with four inocula (10-10(4) colony-forming units) performed in quintuplicate. RESULTS: The combination of Plus Aerobic/F and Plus Anaerobic/F vials detected all the tested pathogenic bacteria, all the tested Gram-positive skin commensal or environmental bacteria, all the tested yeasts, and three of four tested filamentous fungi. The addition of the Mycosis IC/F vial extended the range of detected microorganisms to one fungal environmental contaminant. Two bacterial environmental contaminants were not detected by our method. Low inocula of the skin contaminant Propionibacterium acnes were detected only after 7 days of incubation. CONCLUSIONS: These data suggest that (i) the prolongation of the incubation time of Plus Aerobic/F and Plus Anaerobic/F vials from 7 to 10 days and (ii) the use of Mycosis IC/F medium make minor contributions in the sterility testing of cell therapy products. We have validated the Bactec method using aerobic and anaerobic vials incubated 7 days at 35°C.

Consulter en ligne

Suggestions

Du même auteur

Hospital-diagnosed infections with Escherichia coli clonal group ST131 are mostly acquired in the community

Archive ouverte | Muller, Allison | CCSD

International audience. Abstract The worldwide spread of E. coli ST131 has significantly contributed to the dissemination of E. coli producing extended-spectrum β-lactamases (ESBL). In a French University hospital, ...

Association between Pseudomonas aeruginosa positive water samples and healthcare-associated cases: nine-year study at one university hospital

Archive ouverte | Bouxom, Hélène | CCSD

International audience

Persistent contamination of a hospital hot water network by Legionella pneumophila

Archive ouverte | Jeanvoine, Audrey | CCSD

International audience. We assessed the contamination with Legionella pneumophila (Lp) of the hot water network (HWN) of a hospital, mapped the risk of contamination, and evaluated the relatedness of isolates. We fu...

Chargement des enrichissements...