[A tumour-mimic pig liver model for guiding focused ultrasound thermal ablation]

Archive ouverte

Melodelima, D. | N'Djin, Wa | Parmentier, H. | Chesnais, S. | Rivoire, M. | Chapelon, J.-Y.

Edité par CCSD ; Elsevier Masson -

International audience. There is no established liver tumour model in pigs to study the efficacy of ablative treatment options available for the treatment of liver tumours by physical agents. A tumour-mimic model visible with high contrast on sonograms and on gross pathology has been studied at mid-term on 20 pigs. The aim was to determine if these tumour-mimics are well tolerated and can be used to validate the use of thermal therapies at a preclinical stage. The dimensions of the tumour-mimics measured on sonograms were reproducible (diameter: 9.6 +/- 1.9 mm) and correlated with those performed in gross pathology (R(2)=0.73). The accuracy of focused ultrasound thermal therapy can be evaluated preclinically using these tumour-mimics.

Consulter en ligne

Suggestions

Du même auteur

New integrated imaging high intensity focused ultrasound probe for transrectal prostate cancer treatment.

Archive ouverte | Pichardo, S. | CCSD

International audience. The present study proposes a new integrated imaging (II) high-intensity focused ultrasound (HIFU) probe intended as an improvement to the Ablatherm prostate cancer treatment. Because of a per...

Development of a Simple In Vitro Artery Model and an Evaluation of the Impact of Pulsed Flow on High-Intensity Focused Ultrasound Ablation

Archive ouverte | Cilleros, C. | CCSD

International audience. BackgroundMany parameters have to be investigated before an optimal strategy for thermal therapies can be defined. Such studies are limited by the number of animals that may be included and b...

High-Intensity Focused Ultrasound Using a Toroidal Transducer as an Adjuvant Treatment for Placenta Accreta: A Preliminary Ex Vivo Study

Archive ouverte | Caloone, J. | CCSD

International audience